Related Experiment Video
Updated: Jul 12, 2026

Automated 90Sr Separation and Preconcentration in a Lab-on-Valve System at Ppq Level
Published on: June 6, 2018
Fast determination of 90Sr in urine samples and internal dose evaluation in emergency situations
D Arginelli1, M C Botta1, S Ridone1
1ENEA-CR Saluggia, Radiation Protection Institute, Integrated Laboratory of Radioactivity Measurement and Monitoring, via per Crescentino 41, I-13040, Saluggia, VC, Italy.
Abstract:
During nuclear or radiological emergency, it is important to rapidly identify contaminated people to administer them necessary treatments as soon as possible. A simple method to project and adapt measurement techniques for a mass monitoring of internal contamination based on the concept of "minimum detectable dose" has been developed. This method has been applied to 90Sr, as it has a specific significance for spread in anthropic activity and radiotoxicity, so to require a fast, precise and suitable technique for its determination in many biological samples. A chromatographic separation allowed to isolate and measure by gas proportional counting 90Sr excreted into urine and to calculate the correspondent intake and committed effective dose within half a day, taking also into account an overestimation of 10 % of total activity due to 90Y not eliminated or formed during the measurement. The capacity to attribute values of committed effective dose lower than the precautionary reference level of 10 mSv, which is generally considered safe and does not typically cause health observable effects, has proved the method efficacy, even for a sample collection range larger than 20 days from the intake and for the most dangerous Type S of the investigated radionuclide.