A rapid and sensitive UHPLC-MS/MS method for epalrestat detection in micro-volumes of human plasma for the first time
Hong Chen1, Kewei Liu1, Jiawen Zhang1
1School of Pharmacy, Shenyang Pharmaceutical University, Shenyang, Liaoning Province, P. R. China.
Aim:
A rapid and sensitive ultra-high-performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) method was developed and validated for epalrestat detection in human plasma.
Materials And Methods:
A triple quadrupole tandem mass spectrometer equipped with an electrospray ionization (ESI) source was used to quantify epalrestat and the internal standard epalrestat-d5 in the negative ion mode using multiple reaction monitoring (MRM). After acetonitrile-mediated protein precipitation, chromatographic separation was achieved using a reversed-phase C18 column (ACQUITY UPLC BEH, 2.1 × 50 mm, 1.7 μm; Waters Corp) with acetonitrile and 2 mM ammonium acetate in water as the mobile phase through gradient elution.
Results And Conclusion:
The retention time of epalrestat was 0.92 min and the entire run time was only 2.00 min. The calibration curve was linear in the range 10.0-8.00 × 103 ng/mL (r ≥ 0.99). The within-run and between-run relative standard deviations (RSDs) were < 9.3%. The within-run and between-run relative errors (REs) ranged -8.4-4.2%. No significant matrix effects were observed and the recovery rate was high. The method was fully validated, including reinjection reproducibility in human plasma, and was successfully applied in a pharmacokinetic study, in which 100% incurred sample reanalysis met the criteria.
More Related Videos
08:41Integration of Miniaturized Solid Phase Extraction and LC-MS/MS Detection of 3-Nitrotyrosine in Human Urine for Clinical Applications
Published on: July 14, 2017
14:39Semi-Targeted Ultra-High-Performance Chromatography Coupled to Mass Spectrometry Analysis of Phenolic Metabolites in Plasma of Elderly Adults
Published on: April 22, 2022