Comprehensive Impact of Multiplanar Malalignment on Prosthetic Mechanics Under Gait Loading After Total Knee
Yichao Luan1,2, Min Zhang1, Xiang Dong2,3
1School of Biological Science and Medical Engineering, Beihang University, Beijing, China.
Background:
Component alignment is a key factor influencing clinical outcomes after total knee arthroplasty (TKA). Previous studies have shown that single-plane alignment can significantly affect knee joint kinematics and biomechanics. However, the comprehensive impact of multiplanar malalignment has been rarely investigated.
Objective:
This study aimed to investigate the influence of the multiplanar malalignment combination on the polyethylene tibial liners under gait loading, a primary activity of daily life, as well as the degree of the influence of the alignments on the different planes.
Method:
A validated finite element model of a cruciate-retaining knee prosthesis under gait loading was used in this study. Five alignment parameters (-5°, -3°, 0°, 3°, 5°) on each plane (coronal, sagittal, and transverse) were selected to simulate clinical alignment errors, resulting in 125 models combining various alignment errors across the three planes. Boundary and loading conditions were set according to ISO 14243-3:2014. The maximum von Mises stress and contact stress during a gait cycle were recorded for statistical analysis. A polynomial model was used for regression analysis, with the degree of influence of each alignment error on von Mises and contact stress determined by examining the quadratic coefficients.
Results:
The highest Mises and contact stress values occurred with alignment errors of 5° varus, 5° flexion, and 5° internal rotation on the coronal, sagittal, and transverse planes, respectively. The lowest stress values were observed with a combination of 3° valgus, 5° flexion, and 0° internal rotation. The regression analysis yielded an R 2 value of 0.69 between alignment errors and Mises stress, with quadratic coefficients of 0.096, 0.013, and 0.064 for the coronal, sagittal, and transverse alignments, respectively. For contact stress, the R 2 was 0.697, with quadratic coefficients of 0.083, 0.002, and 0.026 for the coronal, sagittal, and transverse alignments, respectively.
Conclusion:
The coronal alignment of the lower limb has the most significant impact on both Mises stress and contact stress of the tibial liner, followed by the rotational alignment of the tibial component. In contrast, the sagittal alignment of the femoral component has the least influence.
Related Concept Videos
Equations of Equilibrium in Three Dimensions
According to the vector equations of equilibrium, the vector sum of all the external forces acting on a body must...
Deformation of Member under Multiple Loadings
In the case of a member with a variable cross-section, the strain is not constant but depends on the position. The deformation of an...
Plastic Deformations of Members with a Single Plane of Symmetry