Corneal and Epithelial Thickness Measurements: Comparability between two Different Ocular Coherence Tomography
Burcu Yakut1, Anika Förster2, Burkhard Dick3
1Department of Ophthalmology, Haseki Training and Research Hospital, Istanbul, Turkey.
Summary
Optical coherence tomography (OCT) devices provide valuable corneal and epithelial thickness measurements for diagnosing eye conditions. However, measurements vary between devices, meaning they should not be used interchangeably in clinical practice.
Area of Science:
- Ophthalmology
- Medical Imaging
- Corneal Science
Background:
- High-frequency ultrasound and anterior segment optical coherence tomography (OCT) are used for diagnosing keratoconus and planning refractive surgery.
- Anterior segment OCT is increasingly utilized for monitoring keratoconus progression, hydrops, and epithelial healing.
- Comparing different OCT devices is crucial for consistent diagnostic and surgical planning data.
Purpose of the Study:
- To compare epithelial and pachymetric thickness parameters from two different Fourier-domain OCT devices.
- To determine if measurements from two OCT devices can be used interchangeably for corneal and epithelial thickness analysis.
- To assess the interchangeability of OCT measurements in normal and post-SMILE eyes.
Main Methods:
- Retrospective chart review of 24 normal and 32 post-SMILE eyes.
- Corneal and epithelial thickness maps were acquired using two OCT devices: Cirrus 5000 and Avanti RTVue 100.
- Key parameters analyzed included minimum corneal thickness (Pachy min), minimum minus median corneal thickness (Pachy min-med), and maximum epithelial thickness (Epi max).
Main Results:
- Pachy min and Epi max measurements showed good agreement between the two OCT devices (ICC 0.906 and 0.839).
- Significant correlations were found for Pachy min, Pachy min-med, and Epi max in normal eyes.
- In post-SMILE eyes, only Pachy min and Epi max measurements showed significant correlation between devices.
Conclusions:
- Anterior segment OCT is valuable for quantitative corneal and epithelial thickness measurements.
- Significant differences exist in measurements between OCT devices.
- OCT device measurements should not be used interchangeably due to device-specific variations.