Related Experiment Video
Updated: Jun 15, 2025

High-Throughput Analysis of Optical Mapping Data Using ElectroMap
Published on: June 4, 2019
The signal quality of tripolar Laplacian electrogram compared to bipolar electrogram in cardiac electrophysiology
Soichiro Yamashita1, Makoto Takemoto1, Wataru Fujimoto1
1Department of Cardiology Hyogo Prefectural Awaji Medical Center Sumoto Japan.
Background:
The usefulness of tripolar Laplacian electrogram (TLE) in cardiac electrophysiology is unknown.
Objective:
We investigated the difference in the morphology, voltage amplitude, noise level, and far-field potential (FFP) between TLE and bipolar electrogram (BE).
Methods:
Twenty-two patients who underwent catheter ablation for atrial fibrillation were analyzed. TLE and BE were simultaneously obtained using 64-electrode basket catheters during atrial tachycardia or pacing. Local electrograms (EGMs) were analyzed at the locations of three activation patterns: normal conduction, wavefront collision, and slow conduction (SC). In addition, each activation pattern was analyzed at two different wavefront angles: horizontal and vertical, respectively. The voltage amplitude, duration, morphology, and deflection number were compared. The noise level and FFP amplitude were also analyzed.
Results:
With all activation patterns and wavefront angles, the voltage amplitude was significantly lower on TLE than on BE, and the mean amplitude ratio of TLE/BE was 0.75. The duration and deflection number of EGM were comparable. The ventricular FFP amplitude was lower on TLE than on BE (0.04 mV vs. 0.40 mV, p = .01). TLE also had less noise than BE (0.08 mV vs. 0.15 mV, p = .02), with a noise reduction of 47%. At the SC area, low-voltage EGMs below the mean noise levels were more common with BE than with TLE (29% vs. 8%, p = .04).
Conclusion:
TLE has a lower voltage amplitude than BE for all activation patterns and wavefront angles, and it removes more noise and FFPs. TLE is useful for analyzing low-voltage and fractionated EGMs regardless of the wavefront angle.
More Related Videos
Related Concept Videos
Bode Plots Construction
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Electrophysiology of Normal Cardiac Rhythm
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...