Validation of cobas® pulse point-of-care testing device for blood glucose monitoring
Josefine B H Adelhelm1, Charlotte S Jørgensen1, Heidi B Hansen1
1Department of Clinical Biochemistry, Odense University Hospital, Odense, Denmark.
Abstract:
The aim of this study was to validate the blood glucose point-of-care system, cobas® pulse (Roche Diagnostics GmbH), which is the successor to the Accu-Chek® Inform II system (Roche Diagnostics GmbH). Since the cobas® pulse device is intended to replace an existing device from the same manufacturer, we found it highly relevant to perform an industry-independent validation regarding accuracy and comparability with existing glucose measurement systems. From 40 randomly selected, non-fasting adults capillary and venous blood was drawn simultaneously. Correlation and agreement was evaluated by comparing capillary blood glucose on cobas® pulse to plasma glucose on cobas® 8000 (Roche Diagnostics GmbH) and capillary blood glucose on ABL800 Flex (Radiometer, Denmark), respectively. The cobas® pulse generally showed good agreement with both comparison methods, although the agreement between cobas® pulse and ABL800 Flex was better (bias -0.01 mmol/L) than between cobas® pulse and cobas® 8000 (bias 0.61 mmol/L). Differences between measurements of low blood glucose levels (range 0.5 to 4.8 mmol/L) and higher blood glucose levels (range 9.7 to 15.3 mmol/L) when comparing cobas® pulse to ABL800 Flex was also within allowable limits. Altogether, the validation study demonstrated a clinically satisfactory performance of the cobas® pulse point-of-care device.
More Related Videos
06:16Signal Acquisition, Score Interpretation, and Economics of a Non-Invasive Point-of-Care Test for Coronary Artery Disease
Published on: August 9, 2024
13:42Dry Film Photoresist-based Electrochemical Microfluidic Biosensor Platform: Device Fabrication, On-chip Assay Preparation, and System Operation
Published on: September 19, 2017
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Pulse Assessment Sites
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: