Benchmarking test-retest variability in microperimetry for intermediate age-related macular degeneration using MP-3

Jakob Schweighofer1, Klaudia Birner2, Hamza Mohamed3

  • 1Department of Ophthalmology and Optometry, Medical University of Vienna, Vienna, Austria.

Abstract

Insights

Microperimetry (MP) shows good test-retest reliability in intermediate age-related macular degeneration (iAMD). Devices MP-3 and MAIA demonstrate comparable outcomes, supporting MP

Area of Science:

  • Ophthalmology
  • Medical Technology
  • Clinical Trials

Background:

  • Microperimetry (MP) is a key functional endpoint for nonexudative age-related macular degeneration (AMD).
  • Establishing reliable test-retest values is crucial for its clinical application in intermediate AMD (iAMD).

Purpose of the Study:

  • To provide reference values for test-retest outcomes of two MP devices (MP-3 and MAIA) in patients with iAMD.
  • To assess the impact of specific biomarkers on MP test-retest repeatability.

Main Methods:

  • A prospective, cross-sectional study involving 20 subjects with iAMD.
  • Consecutive testing on MP-3 and MAIA devices, with point-wise sensitivity (PWS) superimposed on OCT images.
  • Analysis of biomarkers including hyperreflective foci (HRF), drusen, ellipsoid zone (EZ)-thickness, outer nuclear layer (ONL)-thickness, and subretinal drusenoid deposits (SDD).

Main Results:

  • Both MP-3 and MAIA demonstrated comparable overall test-retest variances (±4.54 dB and ±5.24 dB, respectively).
  • Subretinal drusenoid deposits (SDD) significantly impacted MAIA's repeatability (p=0.03).
  • Good intradevice and interdevice correlations were observed for both devices.

Conclusions:

  • Microperimetry (MP) exhibits good intradevice and interdevice repeatability in iAMD patients using MP-3 and MAIA devices.
  • Biomarkers, excluding SDD, do not significantly affect test-retest repeatability.
  • These findings support MP as a reliable functional endpoint in clinical trials for iAMD.