The digital tourniquet: how safe is it?

J D Lubahn, J Koeneman, K Kosar

    The Journal of Hand Surgery
    |September 1, 1985
    PubMed
    Summary

    This study measured digital tourniquet pressure using latex rubber properties. A new method using a Penrose drain with specific markings can consistently control pressure between 250-375 mm Hg.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Lymphocyte count and NLR as predictive value for the severity of acute cholangitis.

    European review for medical and pharmacological sciences·2023
    Same author

    Comparison of infliximab with adalimumab in biologic-naïve patients with Crohn's disease: a single-center 13-year experience.

    European review for medical and pharmacological sciences·2023
    Same author

    Factors affecting the outcome of the acidification power test of yeast quality: critical reappraisal.

    Folia microbiologica·2007
    Same author

    The anatomical relationship between the posterior interosseous nerve and the supinator muscle.

    The Journal of hand surgery·2000
    Same author

    Orthopaedics, ethics, and industry. Appropriateness of gifts, grants, and awards.

    Clinical orthopaedics and related research·2000
    Same author

    Open-palm technique and soft-tissue coverage in Dupuytren's disease.

    Hand clinics·1999

    Area of Science:

    • Biomedical Engineering
    • Materials Science

    Background:

    • Digital tourniquets are used in medical procedures.
    • Accurate pressure measurement and control are crucial for tourniquet effectiveness and safety.

    Purpose of the Study:

    • To quantify the pressure generated by common digital tourniquets (Penrose drain, rolled glove finger).
    • To develop a method for controlled pressure application using a Penrose drain.

    Main Methods:

    • Mathematical modeling of pressure-resistance in latex rubber.
    • Measurement of pressures beneath rolled glove fingers (200-1200 mm Hg) and Penrose drains (100-650 mm Hg).
    • Development of a technique using marked Penrose drains to achieve controlled pressure (250-375 mm Hg).

    Main Results:

    • Rolled glove fingers generated pressures from 200-1200 mm Hg.
    • Penrose drains produced pressures ranging from 100-650 mm Hg.
    • The developed method with marked Penrose drains achieved consistent pressures between 250-375 mm Hg.

    Conclusions:

    • Digital tourniquet pressures vary significantly depending on the material and application.
    • A novel technique using marked Penrose drains allows for predictable and controlled pressure application.
    • This method offers improved safety and efficacy for digital tourniquet use.

    Related Experiment Videos