Numerical investigation of blood rheology in ventricular assist devices: effects on performance and shear stress
Mohamed Bounouib1, Reda Lakraimi1, Mourad Taha-Janan1
1Laboratory of Applied Mechanics and Technologies, ENSAM, Mohammed V University in Rabat, Morocco.
Abstract:
This study evaluates whether Newtonian models can replace non-Newtonian models in ventricular assist device (VAD) simulations. Five rheological models were compared in an axial-flow VAD using ANSYS CFX. High-shear conditions (92% > 300 s-1 rendered non-Newtonian effects negligible, with errors <1% for pressure rise, efficiency, and torque. Wall shear stress variations were minimal (±5 Pa) and below hemolysis thresholds. Newtonian models suffice for performance predictions in high-shear regions, reducing computational costs by 30-50%. However, localized non-Newtonian effects in stagnation zones may need analysis for thrombogenicity. These findings streamline VAD design without compromising accuracy.