Fluorescence X-ray field for radiation dosemeters calibration
Tomoya Tsuji1, Hiroshi Yoshitomi1, Fuminobu Sato2
1Department of Radiation Protection, Nuclear Science Research Institute, Japan Atomic Energy Agency, 2-4 Shirakata, Tokai-mura, Naka-gun, Ibaraki 319-1195, Japan.
Radiation Protection Dosimetry
|June 4, 2025
Summary
A new fluorescence X-ray calibration field was established for dosemeter calibration. This provides crucial reference air kerma rates, improving radiation measurement accuracy in Japan.
Area of Science:
- Medical Physics
- Radiation Dosimetry
- Metrology
Background:
- The Facility of Radiation Standards at the Japan Atomic Energy Agency re-established a fluorescence X-ray calibration field.
- This field is essential for calibrating dosemeters in a specific X-ray energy range.
- Existing calibration fields in Japan did not cover this specific fluorescence X-ray spectrum.
Purpose of the Study:
- To establish a reliable irradiation environment for dosemeter calibration using fluorescence X-rays.
- To determine reference air kerma rates accurately within this specialized field.
- To derive necessary correction factors for measurements using fluorescence X-rays.
Main Methods:
- Re-establishment of the fluorescence X-ray calibration field.
- Determination of reference air kerma rates using calibrated reference ionization chambers.
- Correction of measurements to account for energy differences between fluorescence X-ray and narrow-spectrum calibration fields.
- Derivation of conversion coefficients and air density correction factors from measured X-ray spectra.
Main Results:
- Successful re-establishment of the fluorescence X-ray calibration field.
- Determination of reference air kerma rates for the new field.
- Calculation of correction factors essential for accurate dosimetry in this specific X-ray environment.
- Generation of data for conversion coefficients and air density corrections, including values beyond the ISO 4037 standard.
Conclusions:
- The re-established fluorescence X-ray calibration field provides a vital standard for dosemeter calibration in Japan.
- Accurate reference air kerma rates and necessary correction factors have been determined.
- This work enhances the accuracy and reliability of radiation measurements for fluorescence X-ray fields, addressing a gap in national metrology standards.
Related Concept Videos
Fluorescence and Phosphorescence: Instrumentation
766
Fluorometers and spectrofluorometers are two types of instruments used for measuring molecular fluorescence. These instruments differ in how they select excitation and emission wavelengths and the type of light sources they utilize. Fluorometers use absorption interference filters to choose excitation and emission wavelengths. The excitation source in a fluorometer is typically a low-pressure mercury vapor lamp that emits intense lines distributed throughout the ultraviolet and visible regions.
766
Atomic Fluorescence Spectroscopy
515
Atomic fluorescence spectroscopy (AFS) is an analytical technique that involves the electronic transitions of atoms in a flame, furnace, or plasma being excited by electromagnetic (EM) radiation. When these atoms absorb energy, they become excited and subsequently release energy as they return to their original state. This emitted light, or "fluorescence," is observed at a right angle to the incident beam. Both absorption and emission processes transpire at distinct wavelengths, which...
515
Flame Photometry: Lab
377
In a flame photometer, when a solution like potassium chloride is aspirated into the flame, the solvent evaporates, leaving behind dehydrated salt. This salt dissociates into free gaseous atoms in their ground state. Some of these atoms absorb energy from the flame, leading to their excitation. The excited atoms return to the ground state, emitting photons at characteristic wavelengths. Because only electronic transitions are involved, the resulting emission lines are very narrow. The intensity...
377
X-ray Imaging
7.8K
German physicist Wilhelm Röntgen (1845–1923) was experimenting with electrical current when he discovered that a mysterious and invisible "ray" would pass through his flesh but leave an outline of his bones on a screen coated with a metal compound. In 1895, Röntgen made the first durable record of the internal parts of a living human: an "X-ray" image (as it came to be called) of his wife’s hand. Scientists worldwide quickly began their own experiments with...
7.8K