Comparison of hemolysis with different pulsed field ablation systems
Iwanari Kawamura1, Shinsuke Miyazaki1, Ryosuke Kato1
1Department of Cardiovascular Medicine, Institute of Science Tokyo, Tokyo, Japan.
Heart Rhythm
|June 5, 2025
Summary
Pulsed field ablation (PFA) systems show varying levels of hemolysis, a known complication. Farapulse demonstrated significantly higher hemolysis biomarkers compared to PulseSelect, indicating potential differences in safety profiles.
Area of Science:
- Cardiology
- Medical Devices
- Electrophysiology
Background:
- Hemolysis is a unique complication associated with pulsed field ablation (PFA).
- Understanding the differences in hemolysis between various PFA systems is crucial for patient safety.
- Current knowledge regarding comparative hemolysis across PFA technologies is limited.
Purpose of the Study:
- To investigate and compare the incidence and severity of hemolysis between two distinct PFA systems: PulseSelect and Farapulse.
- To evaluate hemolysis biomarkers and myocardial injury markers post-procedure.
- To compare PFA systems against cryoballoon ablation as a control.
Main Methods:
- A comparative study involving patients undergoing PFA for atrial fibrillation (PulseSelect: 38 patients, Farapulse: 38 patients) and cryoballoon ablation (14 patients).
- Measurement of free hemoglobin levels pre- and post-procedure.
- Assessment of hemolysis biomarkers (lactate dehydrogenase, total bilirubin, haptoglobin) and myocardial injury biomarkers (creatine kinase) within 24 hours post-ablation.
Main Results:
- No significant difference in the total number of PFA applications between PulseSelect and Farapulse.
- Farapulse group showed significantly higher levels of free hemoglobin, lactate dehydrogenase, and total bilirubin compared to PulseSelect.
- A higher incidence of low haptoglobin levels was observed in the Farapulse group, indicating more pronounced hemolysis.
- No significant difference in hemolysis biomarkers between PulseSelect and cryoballoon ablation.
- Both PFA systems showed higher creatine kinase levels than cryoballoon ablation, but no difference between PFA systems.
- No acute kidney injury was reported in any group.
Conclusions:
- The degree of hemolysis can differ significantly between PFA systems.
- Further research is warranted to establish system-specific safe application limits for PFA to mitigate hemolysis risks.
Related Concept Videos
Blood Flow
Blood is pumped by the heart into the aorta, the largest artery in the body, and then into increasingly smaller arteries, arterioles, and capillaries. The velocity of blood flow decreases with increased cross-sectional blood vessel area. As blood returns to the heart through venules and veins, its velocity increases. The movement of blood is encouraged by smooth muscle in the vessel walls, the movement of skeletal muscle surrounding the vessels, and one-way valves that prevent backflow.
Overview of Blood Vessels
The human cardiovascular system comprises five primary types of blood vessels: arteries, arterioles, veins, venules, and capillaries, each serving unique functions.
Arteries and Arterioles: Arteries are muscular and elastic vessels that primarily carry oxygenated blood from the heart to body tissues, except for the pulmonary artery, which carries deoxygenated blood. They have thick walls to withstand high pressure and contain a layer of muscle tissue, allowing them to expand or contract as...
Arteries and Arterioles: Arteries are muscular and elastic vessels that primarily carry oxygenated blood from the heart to body tissues, except for the pulmonary artery, which carries deoxygenated blood. They have thick walls to withstand high pressure and contain a layer of muscle tissue, allowing them to expand or contract as...
Pulse
When the heart pumps blood out, arterial elastic fibers play a crucial role in sustaining a high-pressure gradient. They expand to accommodate the received blood and then recoil - a process known as the pulse that can be either manually palpated or electronically quantified. Despite a reduction in its effect with increased distance from the heart, elements of the pulse's systolic and diastolic components persist, observable even at the arteriole level.
The pulse serves as a clinical indicator...
The pulse serves as a clinical indicator...
Applications of Integration to Find Blood Flow
Blood flow through a cylindrical blood vessel can be mathematically described using the principles of laminar flow, a regime in which fluid moves smoothly in parallel layers. In this model, the velocity of the blood is not uniform across the cross-section of the vessel; rather, it varies with the radial distance from the center. The maximum velocity occurs along the central axis, decreasing progressively toward the vessel walls, where it reaches zero due to viscous drag.Approximating Blood...