Three-dimensional joint bright, gray, and black blood magnetic resonance imaging technique for multi-parametric
Background:
Multi-parametric imaging of the carotid artery enables quantitative characterization of vulnerable atherosclerotic plaques, which is crucial for preventing ischemic stroke. However, the existing sequential acquisition-based multi-parametric imaging techniques of carotid artery lack bright blood imaging, which is essential for plaque components identification and boundary delineation. This study aims to develop a joint bright, gray, and black blood imaging technique for carotid artery multi-parametric imaging and validate its accuracy and feasibility.
Methods:
The proposed technique incorporated variable flip angles, variable duration of improved motion-sensitized driven equilibrium prepulse, and variable time of echo mapping strategies with three-dimensional multi-shot spoiled gradient-recalled echo acquisition, generating T1, T2, and T2* maps. Bright, gray, and black blood images were sequentially acquired in six scans covering the entire extracranial artery with isotropic resolution (0.7 mm) when natural inflow blood enhancement and imposed blood suppression module were alternatively performed. A B1-specific dictionary was simulated and matched to the measured signal for T1 and T2 estimation while least square fitting was applied for T2* estimation. The proposed technique was compared against reference sequences and validated in healthy volunteers (n = 8) and patients (n = 4) with carotid atherosclerotic plaques.
Results:
The proposed technique achieved an agreement of R2 = 0.99 in T1, T2, and T2* measurements with standard sequences in phantom study. In healthy volunteer study, the proposed technique reached high intraclass correlation coefficients (ICC: 0.906-0.956) with reference sequences in measuring T1, T2, and T2* of cervical muscle, but overestimation and underestimation were observed in T1 (against modified Look-Locker inversion-recovery, bias = 4.8%) and T2 (against gradient and spin echo sequence, bias = -3.3%), respectively. No significant difference was found in the measurement of morphology and quantitative parameters between scan and rescan, while excellent intra- (ICC: 0.804-0.999) and inter-observer (ICC: 0.816-0.982) repeatability was reached. In patient study, the proposed technique demonstrated reliable performance in analyzing vascular morphology and characterizing plaque components with distinctive signal characteristics and quantitative values.
Conclusion:
The proposed technique enables joint bright, gray, and black blood imaging technique for carotid artery multi-parametric imaging with large coverage and isotropic resolution, indicating clinical potential for comprehensive characterization of carotid vulnerable plaque.
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT