Related Experiment Video
Updated: Sep 19, 2025

Radiotracer Administration for High Temporal Resolution Positron Emission Tomography of the Human Brain: Application to FDG-fPET
Published on: October 22, 2019
The use of simulations to study the effect of acquisition schemes and quantification methods using positron emission
Mohlapoli S Mohlapholi1,2, Alex Doruyter1,3, Michael Mix1,4
1Division of Nuclear Medicine, Department of Medical Imaging and Clinical Oncology, Faculty of Medicine and Health Sciences, Stellenbosch University.
Introduction:
This computational study evaluates the accuracy of kinetic models and acquisition schemes in dynamic PET imaging using simulations of 18 F-fallypride PET in the human brain on the real-world data.
Methods:
We employed a 2-tissue 4-k model to generate ideal tissue curves for three regions (putamen, thalamus, and temporal cortex) and a reference region (cerebellum), incorporating a simulated metabolite-corrected input function. Realistic measurements were simulated over a 240-min PET scan by defining acquisition protocols (frame timings and durations), modeling tracer decay, and adding noise. Distribution volume ratios (DVRs) were calculated using the Logan reference analysis and the simplified reference tissue model (SRTM), the relative error in DVR was also assessed across various acquisition protocols. Rate constants from the 2-tissue model were varied, and Bland-Altman analysis was quantified to determine bias relative to ground-truth DVR.
Results:
Results indicate that, under low noise conditions, the Logan reference method performed optimally with a protocol involving a 60-min dynamic scan, a 60-min break, a 30-min scan, another 60-min break, and a final 30-min scan. In noisier conditions, the SRTM yielded the best results with a 150-min effective scan time incorporating three breaks.
Conclusion:
These findings highlight the impact of noise and acquisition strategy on model performance, informing optimal PET imaging protocols.
More Related Videos
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
15:10A Dual Tracer PET-MRI Protocol for the Quantitative Measure of Regional Brain Energy Substrates Uptake in the Rat
Published on: December 28, 2013
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET