Remote Photoplethysmography Technology for Blood Pressure and Hemoglobin Level Assessment in the Preoperative
Selene Y L Tan1, Jia Xin Chai1, Minwoo Choi2
1Division of Anesthesiology and Perioperative Medicine, Singapore General Hospital, Outram Road, Singapore, 169608, Singapore.
Background:
Blood pressure (BP) and hemoglobin concentration measurements are essential components of preoperative anesthetic evaluation. Remote photoplethysmography (rPPG) is an emerging technology that may be used to measure BP and hemoglobin concentration noninvasively with just a consumer-grade smartphone, replacing traditional in-person measurements. However, there is limited data regarding the use of this technology in patients with diverse skin tones and medical comorbidities. Hence, widespread applicability is yet to be achieved. The potential benefits of achieving this would be immense, allowing for greater convenience, accessibility, and reduction in labor and resources.
Objective:
Our study aims to be the first to develop an algorithm for noninvasive rPPG-based BP and hemoglobin concentration measurement that can be used for preoperative evaluation of patients in real-world clinical practice settings.
Methods:
We conducted the study at Singapore General Hospital from March 1, 2023, to June 28, 2024. A total of 200 patients were recruited. Our primary analysis compared the accuracy of rPPG-based systolic and diastolic BP measurements against measurements taken with automated BP measuring devices. Our secondary analysis compared the accuracy of rPPG-based hemoglobin concentration measurement against traditional blood sampling.
Results:
Our model performed best with diastolic BP predictions, with a mean absolute percentage error of 7.52% and a mean difference of 0.16 mm Hg (SD 3.22 mm Hg) between reference and measured readings. The 95% CI for the mean difference between predicted and measured diastolic BP was ±0.57 (-0.41 to 0.73) mm Hg. Systolic BP predictions yielded a mean absolute percentage error of 9.52% and a mean difference of 2.69 mm Hg (SD 7.86 mm Hg). The 95% CI for the mean difference between predicted and measured systolic BP was ±1.14 (-1.54 to -3.83) mm Hg. Hemoglobin concentration predictions had a mean absolute percentage error of 8.52%, with a mean difference of 0.23 g/dL (SD 0.67 g/dL). The 95% CI for the mean difference between predicted and reference measured hemoglobin concentration was ±0.10 (95% CI 0.13-0.33) g/dL.
Conclusions:
Noninvasive rPPG-based measurement of BP and hemoglobin concentration at the preoperative evaluation setting has great potential for improving convenience, improving efficiency, and conserving resources for patients and health care providers. Our model was able to accurately predict diastolic BP in patients with diverse skin tones and medical comorbidities. The findings of this study serve as a basis for further studies to develop and validate the model for noninvasive rPPG-based BP and hemoglobin concentration measurement.
More Related Videos
Related Concept Videos
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)