Enhancing Electrode Location Assessment in Cochlear Implantation via Computed Tomography Image Fusion
Published on: January 17, 2025
Cochlear implant imaging with accelerated flat panel computed tomography: Image quality and dosimetry comparison to
Elie Diamandis1, Sebastian Johannes Müller1, Eya Khadhraoui1
1University Clinic for Neuroradiology, Otto-Von-Guericke-University Magdeburg, Magdeburg, Germany.
None:
Background and PurposeCone-beam CT (CBCT) and high-resolution multislice CT (HR-MSCT) are the mainstay postoperative imaging modalities following cochlear implant (CI) surgery, with CBCT often preferred due to lower susceptibility to metallic artifacts. However, CBCT is more prone to motion artifacts due to longer acquisition times. Recent advancements in accelerated flat panel CT (Acc-FPCT) available with latest generation angiography systems addressed traditional limitations of CBCT by significantly decreasing scan time. This study evaluates the diagnostic performance and radiation dose of Acc-FPCT compared to HR-MSCT in postoperative CI evaluation.MethodsFive cadaveric whole-head specimens (ten temporal bones) were operated on by an ENT surgeon. Ten CIs were inserted via cochleostomy. Post-operatively, specimens were scanned using six Acc-FPCT protocols and HR-MSCT. Three neuroradiologists assessed the image quality of FPCT protocols in comparison to HR-MSCT using a 5-point Likert scale. Seven electrode characteristics including scalar position and discernibility of individual contacts were evaluated. Radiation dose parameters (CTDIvol and DLP) were compared among protocols.ResultsTwo high-resolution Acc-FPCT (HR-FPCT) protocols were rated superior to HR-MSCT (p < .01). There were no significant differences between these two protocols (p = .25). The remaining Acc-FPCT protocols were rated inferior to HR-MSCT (p < .05). Inter-rater reliability was excellent (ICC (2,k) = 0.908; CI [0.85-0.94]). DLP was significantly lower in all Acc-FPCT protocols compared to HR-MSCT.ConclusionsThe results of this study underscore the utility of Acc-FPCT protocols as a feasible alternative to HR-MSCT in postoperative CI evaluation, allowing for better visualization of electrode array while significantly reducing scan duration and radiation exposure.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Design Example: Traverse Angle Computations
Area Computation by the Alternative Coordinate Method
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...