Brain Mapping Using a Graphene Electrode Array
Published on: October 20, 2023
Simulating the Use of Discontinuous Patterned Hydrogel to Improve Inter-Electrode Resistance on Electrode Arrays
Mark L Reeves1, T Jamie Healey1, Avril D McCarthy1,2
1Clinical Engineering, Royal Hallamshire Hospital, Sheffield Teaching Hospitals NHS Foundation Trust, Sheffield, UK.
Background:
A novel form of sensory stimulation aiming to treat spasticity has been developed, and a clinical trial is currently underway. This uses an electrode array controlled by a programmable 64-channel stimulator to spatially vary the electrical stimulation over time. However, when a continuous layer of hydrogel interfaces between the array and skin, stimulation spreads, causing lower current densities applied over larger areas of tissue. A new approach was developed, modeled, and tested, utilizing discontinuous patterned hydrogel to improve inter-electrode resistance on electrode arrays.
Methods:
Finite-difference modeling was used to estimate stimulation distribution within the hydrogel and subcutaneous tissue under the electrode array. Repeated simulations modeled changes due to variations in hydrogel, skin, and subcutaneous tissue resistivity. Properties of both continuous sheets and patterned hydrogel were used for the simulation. Physical prototypes of the continuous and patterned hydrogel were manufactured and tested for comparison with the simulation.
Results:
Simulation results showed a reduced spread of stimulation between electrodes when using the discontinuous patterned hydrogel compared to the continuous hydrogel. This was demonstrated consistently for all variations in hydrogel, skin, and subcutaneous tissue resistivity. Laboratory testing supported the simulation results and suggested the improved performance of the patterned hydrogel, compared with the continuous hydrogel, may become more substantial over time.
Conclusions:
While the simulation only approximates the stimulation distribution on electrode arrays, the results do show potential benefits of utilizing discontinuous patterned hydrogel to increase inter-electrode resistance. Laboratory testing and initial feedback from the clinical trial support the results indicated in the simulations.
Related Concept Videos
Standard Electrode Potentials
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in...
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Potentiometry: Membrane Electrodes
Limits with Oscillating Discontinuities
Improper Integrals: Discontinuous Integrands