Related Experiment Video
Updated: Jul 2, 2026

In Vivo EPR Assessment of pH, pO2, Redox Status, and Concentrations of Phosphate and Glutathione in the Tumor Microenvironment
Published on: March 16, 2018
In vitro measurement of proton RBE: A multi-centric comparison using a harmonized setup
Olga Sokol1, Anders Tobias Frederiksen2,3, Mateusz Sitarz3,4
1GSI Helmholtz Centre for Heavy Ion Research, Darmstadt, Germany.
Background And Purpose:
This study presents a multi-center comparison of in vitro cell survival measurements and RBE calculations following proton irradiations conducted under harmonized experimental conditions across six European institutions participating in the INSPIRE framework.
Materials And Methods:
V79-4 cells were irradiated using spread-out Bragg peak (SOBP) proton fields of two configurations delivering 6 and 8 Gy with widths of 6 and 4 cm, respectively. Each center adhered to a standardized protocol, utilizing the same phantom design to minimize uncertainties related to sample positioning. X-ray reference irradiations were also performed to assess cell radiosensitivity across the participating centers.
Results:
Despite the consistent protocol, significant inter-institutional variability was observed in the survival measurements. For both treatment plans, the largest variation was detected in the most distal points of the SOBP (coefficients of variation of 43 % and 60 % for the 6 Gy and 8 Gy plans, respectively). Kruskal-Wallis statistical test confirmed the significant differences between the centers for each of the measured position in the proton field for both SOBP configurations. Discrepancies were observed in calculated RBE data as well, albeit preserving the expected trend for the values to slightly increase towards the distal edge of the SOBP (up to 1.5 and 1.3 for the 6 Gy and 8 Gy plans, respectively).
Conclusion:
The results of the study highlight the minimal biological variation one could expect performing proton RBE measurements in well-aligned experimental conditions and challenges in conducting large-scale, multi-center radiobiological experiments and inter-comparisons between literature data sets.
More Related Videos
Related Concept Videos
Measurement: Standard Units
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Measurements of Strain