Analysis of the Interference Effects of Ethamsylate on the Determination of Serum Creatinine Levels Using Two Methods
Background:
Accurate quantification of serum creatinine is clinically essential for evaluating renal function. The aim of this study was to investigate the interference of ethamsylate in the detection of serum creatinine via the sarcosine oxidase method and the picric acid method.
Methods:
Thirty fresh serum samples with gradient concentrations of creatinine were collected from patients in our hospital. In accordance with CLSI EP7-A2, paired difference experiments and dose-response experiments were conducted to detect serum creatinine by the sarcosine oxidase method and the picric acid method, respectively. In addition, the interference of ethamsylate on the detection of serum creatinine at different concentrations was observed to compare the interference effect of ethamsylate on the detection of serum creatinine by the sarcosine oxidase method and picric acid method.
Results:
Paired difference tests revealed that 0.3 g/L ethamsylate significantly negatively affected the determination of creatinine at low, medium, and high creatinine concentrations via the sarcosine oxidase method, and the interference values were -53.67 μmol/L, -93.33 μmol/L, and -318.33 μmol/L, respectively. There was positive interference in the determination of low-concentration creatinine by the picric acid method; the interference value was 7.67 μmol/L, but there was no interference in the determination of medium- and high-concentration creatinine. Dose-response experiments revealed that the interference effect of ethamsylate on the determination of creatinine at low, medium, and high concentrations was a cubic regression curve. In the samples with low, medium, and high concentrations of creatinine, the lowest effective doses of ethamsylate interfering with the sarcosine oxidase method were 5.78 mg/L, 6.36 mg/L, and 9.66 mg/L, respectively. The lowest effective doses of ethamsylate that interfered with the low-, medium-, and high-creatinine samples via the picric acid method were 70.16 mg/L, 617.36 mg/L, and 859.19 mg/L, respectively. The interference pattern analysis suggests that the degree of interference diminished as creatinine concentration increased.
Conclusions:
Both the sarcosine oxidase method and the picric acid method are affected by ethamsylate. The picric acid method was relatively less effective and was effective mainly for creatinine samples at low concentrations.
More Related Videos
07:07A High-throughput Method for Measurement of Glomerular Filtration Rate in Conscious Mice
Published on: May 10, 2013
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
Related Concept Videos
Factors Affecting Renal Clearance: Renal Impairment
One condition associated with renal failure is uremia. Uremia is characterized by impaired glomerular filtration and fluid accumulation in the body. This condition hinders the renal clearance of drugs, resulting in drug accumulation and potential...
Serum Studies: Renal Function Tests
Determination of Renal Drug Clearance: Graphical and Midpoint Methods
The graphical method involves plotting the rate of drug excretion in urine against the plasma drug concentration. By analyzing the graph, the clearance can be calculated and obtained. Drugs rapidly excreted by the kidneys exhibit a...
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
Renal Clearance
Renal clearance refers to the volume of plasma cleared of a specific substance, such as creatinine, per unit of time. To measure clearance, urine samples are collected over a 24-hour period during each bladder voiding, followed by a single blood sample at the...
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...