Related Experiment Video
Updated: Jun 12, 2025

A High-Throughput Electrochemiluminescence 7-Plex Assay Simultaneously Screening for Type 1 Diabetes and Multiple Autoimmune Diseases
Published on: May 29, 2020
Detection of Anti-RuvBL1/2 Autoantibodies by Targeted Liquid Chromatography-Tandem Mass Spectrometry
Tom Dehaemers1,2, Birthe Michiels1,2, Rita Derua3
1Department of Laboratory Medicine, University Hospitals Leuven, Leuven, Belgium.
Background:
Anti-RuvBL1/2 autoantibodies are found in patients with systemic sclerosis and systemic sclerosis-myositis overlap syndrome. Anti-RuvBL1/2 antibodies recognize conformational epitopes of the RuvBL1/2 complex, which complicates detection by solid phase assays. Here, we propose a method for detection of anti-RuvBL1/2 autoantibodies based on liquid phase immunoprecipitation combined with quantification of the precipitated RuvBL1/2 by LC-MS/MS.
Methods:
Serum antibodies of patients (n = 13) and controls (n = 60) were captured by and cross-linked to protein A/G magnetic beads and incubated with HeLa nuclear cell extract. Captured proteins were eluted and digested with trypsin. The digested antigen peptides were analyzed by liquid chromatography linked to a triple quadrupole mass spectrometer operating in multiple reaction monitoring (MRM) acquisition mode. Detected signals (peak areas) were quantified. Synthetic proteotypic peptides were used for quality control and quantification purposes.
Results:
The MRM analysis was based on 6 proteotypic peptides (3 for RuvBL1, 3 for RuvBL2) that are unique to each protein. The immunoprecipitation LC-MS/MS MRM approach allowed differentiation between anti-RuvBL1/2 positive and anti-RuvBL1/2 negative samples.
Conclusions:
We propose an immunoprecipitation-based LC-MS/MS MRM method for anti-RuvBL1/2 autoantibody testing. This technology is a promising detection method for autoantibodies targeting conformational epitopes.
Related Concept Videos
Western Blotting
The technique begins with separating proteins from the sample using sodium dodecyl sulfate-polyacrylamide gel electrophoresis (SDS-PAGE), followed by protein transfer, immunoblotting, and finally, protein detection.
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...