Validation of a reverse phase high-performance liquid chromatography method for the detection of major components and
Lin Qu1, Jie Lin1, Benkai Qin1
1Fuan Pharmaceutical Group Ningbo Team Pharmaceutical Co., Ltd, Ningbo, China.
Abstract:
A simple, feasible, isocratic elution and stable reversed-phase high performance liquid chromatography was established and verified. Chromatographic conditions were XY-C18, 4.6x150mm, 5μm column; column temperature 50oC. Mobile phase was (A) 0.025 mol/L sodium perchlorate solution, added 1ml triethylamine, and adjust the pH value to 2.0 with perchloric acid solution; (B) acetonitrile-methanol (70:30). The elution method was gradient elution; The flow rate was 1.0mL/min; the injection volume was 10μL and the wavelengths were 239nm and 258nm. The developed method had been validated according to the ICH guidelines and the system suitability, specificity, LOD, LOQ, linearity, range, accuracy, precision, durability and solution stability of the proposed method had been verified. The relationship between concentration and peak area was linear in the concentration range, with an R2>0.995. The linear relationship was good in the range of impurity concentration and the R2 was more than 0.999. The accuracy and repeatability meet the specified criteria. The method had good durability.
More Related Videos
08:54Determining Four Components in a Lipid Nanoparticle RNA Delivery System by Liquid Chromatography Combined with Evaporative Light Scattering Detector
Published on: May 30, 2025
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Related Concept Videos
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Types of Detectors