Exposure control at the first dark-field chest radiography demonstrator system
M E Lochschmidt1, M Frank1, K Willer1
1Chair of Biomedical Physics, Department of Physics, TUM School of Natural Sciences, Technical University of Munich, 85748 Garching, Germany; Munich Institute of Biomedical Engineering, Technical University of Munich, 85748 Garching, Germany; Institute for Diagnostic and Interventional Radiology, School of Medicine and Health, TUM Klinikum, Technical University of Munich (TUM), 81675 München, Germany.
Purpose:
A first demonstrator system for clinical dark-field chest radiography was recently constructed. Due to its scanning acquisition, conventional exposure regulation systems are not possible. This study presents the adaptation of exposure to each patient individually, using a conventional radiograph as a reference. Additionally, an alternative approach for individual exposure planning based on the patient's body mass index (BMI) is proposed.
Methods:
Using exposure settings from a conventional radiography system with automatic exposure control (AEC), a patient-specific equivalent attenuator thickness was calculated with polyoxymethylene (POM) as a surrogate material. The tube current at the dark-field system was adapted to reach the target detector dose based on this thickness. This approach was verified with both phantom measurements and patient examinations. The correlation between patient body parameters (weight, BMI, upper bust girth, and under-bust girth) and desired tube current was also evaluated.
Results:
A calibration curve was found to transfer individual exposure settings from a conventional device to the dark-field system. Achieved detector doses for both phantom and patient examinations were within permitted ranges. A strong correlation was found between each body parameter and the desired tube current, with BMI showing the strongest correlation (r=0.87,p<10-29).
Conclusions:
An exposure planning approach was successfully implemented at the first clinical dark-field chest radiography system, delivering a detector dose within the required range for different patient sizes. A conventional radiograph with AEC is necessary for this implementation. The strong correlation between BMI and tube current offers an alternative approach for patient-specific exposure control.
More Related Videos
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...