Automated Midline Shift and Intracranial Pressure Estimation based on Brain CT Images
Published on: April 13, 2013
Improved Beam-hardening Reduction in CT - Impact on Image Quality and Diagnostic Certainty in Emergency Imaging of
Foroud Aghapour Zangeneh1, Gonçalo G Almeida1, Susanne Bensler1
1Department of Radiology, Kantonsspital Baden, affiliated Hospital for Research and Teaching of the Faculty of Medicine of the University of Zurich, Baden, Switzerland (F.A.Z., G.D.A., S.B., R.O.V., J.A., T.N., A.E.).
Rationale And Objectives:
To assess the impact of an optimized beam-hardening artifact reduction algorithm on image quality and diagnostic certainty in emergency unenhanced brain CT.
Materials And Methods:
Retrospective study of consecutive patients referred for unenhanced brain CT to rule out traumatic brain injuries in 2023. Images were reconstructed using both (A) a standard and (B) an optimized iterative beam-hardening correction algorithm (iBHC). CT attenuation, image noise and SNR were measured in the cortex of supratentorial and infratentorial regions and in the pons. Posterior Fossa Artifact Index (PFAI) and Subcalvarial Artifact Index (SAI) were calculated. Two neuroradiologists and two emergency radiologists independently compared artifacts and diagnostic certainty between both algorithms using 5-point Likert scales and side-by-side comparisons. A paired Wilcoxon-test with correction for multiple testing was used.
Results:
100 consecutive patients (55 women; 64.1 ± 20 years) were included. CT attenuation was significantly lower for B (all P <.0001). SNR was significantly lower supratentorial (frontal region: 10.5 vs. 12.9, P<.0001) and significantly higher in the pons (5.9 vs. 5.5, P <.0001) for B. PFAI was significantly reduced for B (5.5 vs. 6.4, P <.0001), while there was no significant difference in SAI (P = 0.304). The optimized algorithm was selected as superior in 100%, 100%, 99%, 99% of supratentorial and in 100%, 99%, 99%, 86% of infratentorial cases for readers 1 to 4, respectively.
Conclusion:
An optimized iBHC algorithm demonstrated significantly improved image quality, reduced artifacts and improved diagnostic certainty in emergency unenhanced brain CT.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
09:21Optimized Management of Endovascular Treatment for Acute Ischemic Stroke
Published on: January 18, 2018
Related Concept Videos
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Magnetic Resonance Imaging
Radiological Investigation I: X-ray and CT