Radiation Dose Reduction in Cancer Imaging with New-Model CT Scanners: A Quality of Care Evaluation
Stefania Rizzo1,2, Luca Bellesi3, Ebticem Ben Khalifa1
1Clinic of Radiology, Imaging Institute of Southern Switzerland (IIMSI), Ente Ospedaliero Cantonale (EOC), 6900 Lugano, Switzerland.
Background/Objectives:
The primary aim of this study was to evaluate whether the replacement of roughly one-decade-old computed tomography (CT) scanners with new-model CT scanners were associated with an additional reduction in the radiation dose delivered to oncological patients, in a radiological setting where the optimization of protocols had already reached very low radiation doses. An exploratory secondary objective was to evaluate the potential differences in the objective image quality between the CT scans obtained before and after the installation of the new-generation CT scanners.
Methods:
Chest and abdominal CT examinations conducted for oncologic purposes were retrospectively selected from two time periods-prior to scanner replacement (2022) and following an upgrade (2024)-after five CT systems in our radiology department were replaced. We extracted and compared the CT dose index (CTDI) and dose length product (DLP) for each CT phase. For the objective image quality evaluation, we calculated the signal-to-noise ratio (SNR) and the contrast-to-noise ratio (CNR) at the center of the liver and the aorta. An appropriate statistical analysis was performed and a p-value < 0.05 was considered significant.
Results:
We included 14,601 CT acquisitions, of which 9013 (61.7%) were performed before and 5588 (38.3%) after the replacement of the CT scanners. There were significantly lower values for the CTDI and DLP with the new CT scanners compared to the old ones. The CTDI with the new CT scanners was significantly lower in all phases (p-value = 0.002 for unenhanced phase, and p < 0.001 for arterial, portal venous, and delayed phases). The DLP using the new CT scanners was significantly lower in the arterial, portal venous, and delayed phases (p < 0.001), and it was not significantly different in the unenhanced phase (p = 0.36). There was no significant difference in the SNR at the liver level (p = 0.72) or at the aorta level (p = 0.51). There was no significant difference in the CNR at the liver level (p = 0.24), whereas the CNR was higher with the new CT scanners at the aorta level (p = 0.03).
Conclusions:
The transition to new-model CT scanners resulted in a significant reduction in the radiation dose delivered by chest and abdomen CT scans, without compromising the objective image quality.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
X-ray Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...