Non-Invasive Hemodynamic Monitoring in Transcatheter Aortic Valve Implantation

Thorald Stolte1, Janarthanan Sathananthan2, Jakob Johannes Reichl1

  • 1Department of Cardiology and Cardiovascular Research Institute Basel (CRIB), University Hospital Basel, University of Basel, 4031 Basel, Switzerland.

PubMed

Insights

The ClearSight® system, a non-invasive method, showed lower hemodynamic values and poor correlation compared to invasive monitoring during Transcatheter Aortic Valve Implantation (TAVI). This suggests limitations in its current use for TAVI hemodynamic assessment.

Area of Science:

  • Cardiovascular Medicine
  • Medical Devices
  • Hemodynamics

Background:

  • Aortic stenosis (AS) is common in elderly patients, often treated with Transcatheter Aortic Valve Implantation (TAVI).
  • Current hemodynamic monitoring during TAVI is invasive.
  • The ClearSight® Finger Cuff system offers a non-invasive alternative.

Purpose of the Study:

  • To compare the reliability and feasibility of non-invasive hemodynamic monitoring (ClearSight®) with traditional invasive methods during TAVI.

Main Methods:

  • Prospective observational study of 50 patients undergoing elective TAVI.
  • Invasive hemodynamics via arterial/pigtail catheters; non-invasive via ClearSight®.
  • Data collected on baseline, procedural, and 30-day outcomes.

Main Results:

  • Non-invasive CO, CI, and SV were consistently lower than invasive measurements (p<0.05 for CO and CI).
  • Non-invasive blood pressure was lower than invasive readings.
  • Low correlation (Mean percentage error 52%) between non-invasive and invasive measurements.

Conclusions:

  • The ClearSight® system yielded lower absolute hemodynamic values compared to invasive methods.
  • Low correlation was observed between non-invasive and invasive measurements pre- and post-TAVI.
  • Current non-invasive monitoring may not be reliable for TAVI hemodynamic assessment.