A 24-m3 direct heat-sink calorimeter with on-line data acquisition, processing, and control
The American Journal of Physiology
|October 1, 1985
Summary
A new 24-m3 direct heat-sink calorimeter accurately measures human evaporative and sensible heat loss. Real-time processing ensures stability and rapid response for continuous, reliable human thermoregulation studies.
Area of Science:
- Human thermoregulation
- Calorimetry
- Physiological measurement
Background:
- Accurate measurement of human heat loss is crucial for understanding thermoregulation.
- Previous calorimeters faced challenges with large volumes, response times, and stability.
- Direct heat-sink calorimeters offer potential for precise, continuous measurement.
Purpose of the Study:
- To construct and evaluate a large-volume (24-m3) direct heat-sink calorimeter.
- To assess its performance for continuous measurement of human evaporative and sensible heat loss.
- To validate its accuracy, stability, and response time for long-term human studies.
Main Methods:
- Construction of a 24-m3 direct heat-sink calorimeter.
- Implementation of real-time processing to compensate for physical time constants and delays.
- Calibration and performance testing for linearity, precision, stability, and response time.
- Conducting 25-hour continuous experiments with human subjects.
Main Results:
- The calorimeter demonstrated linear responses for evaporative (0-320 W) and sensible (0-280 W) heat loss with high precision.
- Excellent stability was achieved, with drift <0.6 W and low standard deviations over 24-72 hours.
- Fast response times (95% in 15 min) and independence from calibration source position were confirmed.
- No measurable cross-talk between heat loss types and negligible humidity dependency were observed.
- Human experiments showed no recorded discomfort during 25-hour trials.
Conclusions:
- The 24-m3 direct heat-sink calorimeter is a reliable instrument for continuous human heat loss measurement.
- Real-time processing effectively addressed stability and response-time issues in large-volume calorimeters.
- The calorimeter's performance characteristics support its use in extended human thermoregulation research.