Image reconstruction and analysis of atherosclerosis imaging by [18F]FDG PETCT using digital PET technology
Stephen Rankin1, A J Gemmell2,3, J McClure1
1BHF Glasgow Cardiovascular Research Centre, School of Cardiovascular and Metabolic Health, University of Glasgow, Glasgow, UK.
Purpose:
Digital PETCT scanners have improved spatial resolution and sensitivity. This may have implications for reconstruction parameters and atherosclerosis assessment using [18F]FDG. On a contemporary digital scanner, we assessed European Association of Nuclear Medicine's (EANM)-recommended reconstruction parameters and blood pool methods, compared with a locally-optimised protocol using fewer iterations and subsets than recommended by EANM.
Methods:
Using clinical PETCT and phantom analysis, we quantitatively assessed two reconstructions ('EANM' and 'local') for atherosclerotic assessment using mean contrast recovery (MCR), absolute error and coefficient of variation (CoV). We assessed the impact of each reconstruction on tissue-to-background ratio (TBR). We also assessed the differences within four blood pool regions on repeated imaging over 24-weeks.
Results:
EANM reconstruction yielded higher TBRmax, 4.28 ± 0.65 vs 1.81 ± 0.24 p < 0.0001, than local reconstruction. Phantom data demonstrated a higher RCmax curve with EANM vs local reconstruction. EANM MCR was 1.87 vs 1.23 with local reconstruction, with higher absolute error (2.23 vs 0.61) and variation (7.63% vs 4.14%), vs local reconstruction. Superior vena cava (SVC) offered the most reproducible and reliable blood pool data. Internal jugular vein had lower FDG-uptake compared to other regions, resulting in higher TBRmax, but was less reproducible between scans over 24 weeks.
Conclusion:
Local reconstruction, with fewer iterations and subsets compared to EANM recommendations, resulted in more accurate atherosclerotic assessment on a contemporary digital scanner. Metrics for assessing reconstructions, such as absolute error and CoV, provided valuable information. The blood pool region used affects TBR. SVC appears to provide the most reliable blood pool region.
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
![Quantification of Atherosclerotic Plaque Activity and Vascular Inflammation using [18-F] Fluorodeoxyglucose Positron Emission Tomography/Computed Tomography FDG-PET/CT](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F3777.jpg&w=3840&q=50)