Real-time changes in pulse pressure during a 10-second end-expiratory occlusion test reliably detect preload
Francesco Gavelli1,2, Nello De Vita3,4, Christopher Lai3
1Service de médecine intensive-réanimation, Hôpital de Bicêtre, DMU CORREVE, Université Paris-Saclay, AP-HP, Inserm UMR S_999, FHU SEPSIS, CARMAS, Le Kremlin-Bicêtre, France. francesco.gavelli@uniupo.it.
Background:
The end-expiratory occlusion (EEO) test detects preload responsiveness through changes in cardiac index (ΔCI) during a 15-second respiratory hold at end-expiration. We investigated the diagnostic accuracy of EEO-induced changes in arterial pulse pressure (∆PP), especially when the duration of EEO is reduced to 10'' and 5'', and whether adding an end-inspiratory occlusion (EIO) improves this diagnostic accuracy.
Methods:
In 143 mechanically ventilated patients with sinus rhythm, EEO and EIO were performed while recording ΔCI and ∆PP values. Either a fluid bolus-induced ΔCI ≥ 15% or a passive leg raising-induced ΔCI ≥ 10% defined preload responsiveness. The effects of the EEO and EIO tests on PP and CI were evaluated as the percentage difference between values averaged either over the last five seconds of the 15-sec respiratory holds (ΔPPEEO-15'' and ΔPPEIO-15'', ΔCIEEO-15'' and ΔCIEIO-15''), or between the 5th and the 10th seconds of the 15-sec respiratory holds (ΔPPEEO-10'' and ΔPPEIO-10'', ΔCIEEO-10'' and ΔCIEIO-10''), or during the five first seconds of respiratory holds (ΔPPEEO-5'' and ΔPPEIO-5'', ΔCIEEO-5'' and ΔCIEIO-5'') and baseline.
Results:
Sixty-one (43%) patients were preload responders. Both ∆CIEEO-15'' and ∆CIEEO-10'' were higher in responders than in non-responders' (5.8 [4.5-7.3]% vs. 1.1 [0.1-3.4]% and 3.0 [2.4-4.3]% vs. 0.6 [0.1-1.6]%, respectively; p < 0.001), whereas ∆CIEEO-5'' did not differ between responders and non-responders. ∆PPEEO-5'', ∆PPEEO-10'' and ∆PPEEO-15'' were significantly higher in responders than in non-responders (5.2 [2.8-8.7]% vs. 1.2 [0.3-2.8]%, 7.7 [5.0-12.4]% vs. 1.8 [0.5-3.1]% and 8.1 [5.1-11.8]% vs. 1.5 [0.5-3.0]%, respectively; p < 0.001). For detecting preload responsiveness, compared to the area under the receiver operating characteristic (AUROC) of ∆CIEEO-15'' (0.935 [0.881-0.969]), the AUROC of ∆CIEEO-10'' was similar (0.910 [0.851-0.951], p = 0.410), but the AUROC of ∆CIEEO-5'' was smaller (0.541 [0.456-0.625], p < 0.001); the AUROC of ∆PPEEO-15'' (0.913 [0.857-0.952], p = 0.346), and ∆PPEEO-10'' (0.912 [0.860-0.947], p = 0.336) were similar, but the AUROC of ∆PPEEO-5'' (0.834 (0.763-0.891, p = 0.005) was significantly smaller. Evaluation of ∆CIEEO+EIO and ∆PPEEO+EIO did not enhance reliability of the test at each test duration.
Conclusion:
In ventilated patients with sinus rhythm, real-time changes in PP during a 10-second EEO reliably detect preload responsiveness.
Trial Registration:
No. IDRCB 2010A0095942. Registered 04 January 2010.
Related Concept Videos
Measurement of Blood Pressure
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Pulse
The pulse serves as a clinical...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pulse amplitude and quality
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...