Expiratory pressures generated by heat moisture exchangers in laryngectomized patients
Toshihiko Yamashita1, Hirotaka Shinomiya1, Ken-Ichi Nibu1
1Department of Otolaryngology-Head and Neck Surgery, Kobe University Hospital, Kobe 650-0017, Japan.
Objective:
HMEs (Heat and Moisture Exchangers) are used postoperatively in laryngectomized patients to replace upper airway functions such as heating, humidification, filtration and resistance that are lost by the surgery. The clinical value of HME resistance function is controversial. There are no studies reporting actual measurements of air pressure generated by HMEs during breathing. In this paper we evaluated air pressures generated by HMEs (XtraFlowⓇ, XtraMoistⓇ, LifeGoⓇ, and LifeHomeⓇ. Atos Medical, Japan), compared the obtained data with theoretical pressure, and discussed the comparison with the expiratory pressure of pursed lips breathing reported in the past.
Methods:
Nine patients who had laryngectomy were enrolled in this study. All participants were male, and the mean age was 72.3 ± 6.7. A pressure measurement system using a U-shaped water manometer was created to measure pressures generated by HMEs. Expiratory pressures were measured in all HMEs at rest and in XtraFlowⓇ and LifeGoⓇ during light exercise.
Results:
Peak expiratory pressure was 5.67 mmH2O (95 %CI, 3.24 to 8.11) and time averaged expiratory pressure was 3.00 mmH2O (95 %CI, 1.56 to 4.45). These pressures were consistent with theoretical values. The finer HME, the higher pressure. The coarse filters had lower expiratory pressure but did not exceed the pressure from the fine filters even with the added pressure from the treadmill load (3Mets).
Conclusion:
Endotracheal pressure was approximately less than 10 mmH2O during expiration in Atos HMEs. The fine filters, such as XtraMoistⓇ and LifeHomeⓇ, had greater exhalation pressures than coarse filters such as XtraFlowⓇ and LifeGoⓇ. The pressure drops of HMEs seemed lower than the pressure drops added by pursed lips expiration and HMEs may not be a substitute for pursed lips expiration.
More Related Videos
09:31Normothermic Negative Pressure Ventilation Ex Situ Lung Perfusion: Evaluation of Lung Function and Metabolism
Published on: February 14, 2022
06:22Surfactant Depletion Combined with Injurious Ventilation Results in a Reproducible Model of the Acute Respiratory Distress Syndrome ARDS
Published on: April 7, 2021
Related Concept Videos
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...
Pulmonary Cycle: Exhalation
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Physical Assessment of the Respiratory Tract IV: Auscultation
Breath Sounds
Breath sounds are categorized into vesicular, bronchovesicular, and bronchial.