Comparison between inductively coupled plasma-mass spectrometry and benchtop X-ray fluorescence performance for trace
Kolawole E Adesina1, Stefano A Parducci1, Joseph D Brain2
1School of Health Sciences, Purdue University, West-Lafayette, IN 47906, USA.
Background:
Trace elemental toxicants induce health detriment in almost every organ system in the human body and account for a large amount of environmental and ecological environmental pollution. Traditionally, inductively coupled plasma mass spectrometry (ICP-MS) has been the gold standard for measuring elemental concentrations in biological tissues collected from toxicological and epidemiological studies. However, ICP-MS is often limited by its complexity, cost, and time-intensive nature.
Methods:
This study investigates the feasibility of benchtop X-ray fluorescence (XRF) as an efficient alternative for trace elemental analysis in rat tissues, offering comparable quantification capabilities with enhanced operational simplicity. We conducted a comparative analysis using tissue samples from multiple rat organs, including stomach, eyes, and liver.
Results:
The elemental concentrations of Arsenic (As), Cadmium (Cd), Copper (Cu), Manganese (Mn), and Zinc (Zn) were measured using both ICP-MS and a high-powered benchtop XRF (Epsilon 4, Malvern Panalytical). Our findings demonstrated strong linear regression correlations between the two methods: As (R2 = 0.86), Cd (R2 = 0.81), Cu (R2 = 0.77), Mn (R2 = 0.88), and Zn (R2 = 0.74). The overall Pearson correlation coefficient was r = 0.95 (p ≤ 0.05), indicating high concordance between the mean concentrations obtained from ICP-MS and benchtop XRF. The median minimum detection limits for the elements were 0.12 μg/g, with specific limits for Cd (0.0042 μg/g), Cu (0.040 μg/g), Zn (0.12 μg/g), As (0.25 μg/g), and Mn (0.35 μg/g) over a 7.5-minute measurement period. Bland-Altman analysis revealed high agreement between the two methods, particularly for As, Cu, and Mn.
Conclusion:
These results suggest that both ICP-MS and benchtop XRF are viable for elemental quantification in organ tissues, with benchtop XRF being more practical for low-mass samples. This study shows benchtop XRF's potential for high-throughput, accurate trace element analysis in biological samples, broadening its use in environmental and toxicological research.
Synopsis:
Human and ecological tissues of varying compositions and densities can be measured effectively using benchtop X-ray fluorescence.
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Atomic Emission Spectroscopy: Overview
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Atomic Emission Spectroscopy: Instrumentation