Is there a sonographer effect? Sonographer as a source of variability for Shear Wave Elastography
Elleen Kim1, Ghada Issa2, Alyssa Berube3
1Warren Alpert Medical School, Brown University, Providence.
Sonographer differences introduce significant variability in Shear Wave Elastography (SWE) measurements. Standardizing protocols is crucial to ensure the reliability of SWE values and clinical thresholds.
Area of Science:
- Medical Imaging
- Diagnostic Ultrasound
Background:
- Shear Wave Elastography (SWE) is a non-invasive ultrasound technique used to assess tissue stiffness.
- Variability in SWE measurements can impact diagnostic accuracy and clinical decision-making.
Purpose of the Study:
- To quantify the contribution of sonographer variability to systematic variance in SWE measurements.
- To assess the impact of inter-sonographer differences on the reliability of SWE values.
Main Methods:
- Two studies employed a block design with six trained sonographers scanning six healthy liver volunteers.
- Standardized procedures and the same ultrasound machine were used for all measurements.
- Each sonographer performed ten SWE measurements per volunteer in the right liver lobe.
Main Results:
- Intraclass correlation coefficients (ICC) varied significantly based on scanning day (0.23 vs. 0.83).
- Sonographer differences accounted for 17% of the total variability in SWE measurements.
- Statistically significant differences in SWE values were observed between sonographers, approaching clinical thresholds.
Conclusions:
- Sonographer performance is a significant factor influencing SWE values.
- Standardization of SWE protocols is essential to minimize inter-sonographer variability.
- The inherent variability may challenge the clinical utility of established SWE thresholds.
More Related Videos
09:02Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
11:50High-frequency High-resolution Echocardiography: First Evidence on Non-invasive Repeated Measure of Myocardial Strain, Contractility, and Mitral Regurgitation in the Ischemia-reperfused Murine Heart
Published on: July 9, 2010
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Elastic Strain Energy for Shearing Stresses
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Imaging Studies II: Ultrasonography