Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Quantification of the difference in hounsfield units of an electron density phantom between a conventional and
Catherine Beck1, Ashleigh V Morrice-West2, Peter Muir3
1Melbourne Veterinary School, Faculty of Science, University of Melbourne, Melbourne, VIC, Australia. cbeck@unimelb.edu.au.
Abstract:
In human CT imaging, opportunistic calculation of bone mineral density (BMD) is used to screen for osteoporosis, where BMD is estimated by Hounsfield Units (HU). There is potential to use HU for estimation of BMD in veterinary imaging. However, HU values can vary between different machines and over time. We undertook a serial measurements study to determine (1) the extent of the difference of the mean HU and noise (HU standard deviation) of eight regions of interest (ROI) of an electron density phantom scanned on two CT machines, a Siemens CT (conventional) machine and the Equina by Asto CT (standing) machine; and (2) to determine if the HU values obtained on each machine differed over time. The mean HU from 30 scans of an electron density phantom were generated from each of the two CT machines (N = 60; conventional and standing). Differences between machines were assessed using a linear mixed effects model. The mean HU of each positive value ROI from the conventional machine was decreased compared with the standing machine (P < 0.001) and the difference was greatest for the denser phantom inserts. A calibration curve equation utilising the four bone-equivalent insert plugs of the phantom, with known calcium hydroxyapatite (CaHA) concentration, was generated using linear regression to represent CaHA, and therefore BMD, for the conventional machine (CaHA = -29.58 + 0.98xROImean) and the standing machine (CaHA = -54.53 + 0.95xROImean). For opportunistic calculation of BMD from HU in horses, the differences in HU observed between different CT machines warrants calibration of HU using a bone equivalent phantom.
More Related Videos
07:33Non-invasive Skeletal Muscle Quantification in Small Animals Using Micro-computed Tomography
Published on: November 8, 2024
09:21Human Brown Adipose Tissue Depots Automatically Segmented by Positron Emission Tomography/Computed Tomography and Registered Magnetic Resonance Images
Published on: February 18, 2015
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography