Case report: Masked premature ventricular contractions and Intradevice interaction causing ventricular Arrythmias.
Aravinthan Vignarajah1, Jose Aguilera2, Justin Z Lee2
1Department of Internal Medicine, Cleveland Clinic Fairview Hospital, Cleveland, OH, USA.
Implantable cardioverter-defibrillators (ICDs) can rarely cause arrhythmias. In a cardiac sarcoidosis patient, a specific timing interaction masked premature ventricular contractions, triggering ventricular tachycardia and fibrillation. Reprogramming the ICD resolved the issue.
Area of Science:
- Cardiology
- Electrophysiology
- Medical Devices
Background:
- Implantable cardioverter-defibrillators (ICDs) are crucial for managing life-threatening ventricular arrhythmias.
- Rarely, ICDs can paradoxically precipitate arrhythmias due to device-patient interactions.
- Cardiac sarcoidosis is an inflammatory condition that can affect the heart's electrical system.
Purpose of the Study:
- To report a rare case of ICD-induced polymorphic ventricular tachycardia and ventricular fibrillation.
- To elucidate the specific intradevice timing mechanism leading to arrhythmia precipitation.
- To emphasize the importance of tailored ICD programming in preventing device-related complications.
Main Methods:
- Case report of a 44-year-old male patient with cardiac sarcoidosis.
- Analysis of the interaction between premature ventricular contractions (PVCs) and the ICD's ventricular blanking period.
- Device reprogramming from DDD to VVI mode as an intervention.
Main Results:
- The patient experienced recurrent polymorphic ventricular tachycardia (VT) and ventricular fibrillation (VF).
- PVCs were masked by the ICD's post-atrial pacing ventricular blanking period.
- This "R-on-T" phenomenon, triggered by ventricular pacing during the vulnerable period, induced VT/VF.
- Reprogramming to VVI mode successfully eliminated device-induced arrhythmias.
Conclusions:
- Individualized ICD programming is essential to prevent device-induced arrhythmias.
- Understanding intradevice timing interactions is critical for managing patients with complex cardiac conditions.
- This case underscores the need for careful evaluation of ICD behavior in patients prone to arrhythmias.
Related Concept Videos
Mechanism of Cardiac Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias II: Classification of Tachyarrhythmias
Dysrhythmias V: Evaluating Dysrhythmias
Dysrhythmias VI: Management of Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias