Performance verification of the new UF-1500 urine particle analyser: a new opportunity for small and medium
Giulia Previtali1, Michela Seghezzi1, Roberto Marozzi1
1Clinical Chemistry Laboratory, Department of Laboratory Medicine, Papa Giovanni XXIII Hospital, Bergamo, Italy.
Background:
The UF-1500 is the new fully automated urine particle analyser by Sysmex specifically tailored for small and medium laboratories. This study aim to validate its analytical and diagnostic performance.
Methods:
754 first morning mid-stream urines were analysed on UF-1500; 550 samples were used for the UF-5000 comparison and 204 were used for the correlation with manual count on Fuchs-Rosenthal chamber. Carry-over, linearity and imprecision of the UF-1500 were also assessed.
Results:
Correlation with the UF-5000 was excellent, with r coefficient range between 0.88 and 1.00. Correlation with Fuchs-Rosenthal chamber was very good for all the parameters; r coefficient ranged between 0.67 and 0.94. Linearity regression coefficient of determination (R2) was excellent for almost all the parameters. No carry-over was observed. The within-run imprecision range between 2.93 % for RBC and 35.63 % for WBC. The between-run imprecision ranged between 2.1 % for RBC and 23.9 % for CAST, using low and high positive quality controls, respectively.
Conclusion:
The analytical and diagnostic performance is satisfactory for almost all the parameters, when compared with the UF-5000; the correlation with the reference method is equally good.