World's First Artificial Intelligence-Based Evaluation of Rist Catheter Stability in Transradial Procedures: A
Shunsuke Tanoue1,2, Yuya Sakakura3, Kenichi Kono4,5
1Department of Neurosurgery, Mishuku Hospital, Tokyo, Japan.
Objective:
Artificial intelligence (AI) holds promise for advancing neuroendovascular therapy through device evaluation, but its application in real-world clinical settings remains limited. We aimed to validate the feasibility of AI-based quantitative device evaluation during actual procedures by assessing the stability of the Rist radial access guide catheter (Medtronic, Dublin, Ireland), a novel device designed for the increasingly adopted transradial approach (TRA), during flow diverter stent (FDS) placement.
Methods:
We retrospectively analyzed 4 cases of FDS placement using Rist via the TRA. Rist was tracked in recorded fluoroscopic videos using the AI technology of Neuro-Vascular Assist (iMed Technologies, Tokyo, Japan). The movement distance of Rist during FDS placement was calculated as a stability indicator.
Results:
All procedures were successfully completed without any complications. Rist was introduced from the radial artery and positioned in the distal internal carotid artery. The maximum movement distances of the Rist during the procedures were 3.36, 6.63, 1.79, and 0.33 mm for each case, respectively, with an average of 3.03 mm. The maximum movement distances per minute were 1.68, 2.34, 1.19, and 0.46 mm/min, respectively, with a mean of 1.42 mm/min. These measurements suggest sufficient stability for the FDS procedures.
Conclusion:
This study demonstrates the feasibility of using AI technology to quantitatively analyze Rist stability in TRA procedures. To the best of our knowledge, this is the 1st clinical evaluation of device function in a clinical setting using AI technology. Further studies with more cases are required to validate these findings. This method is promising for real-world device evaluation and development.
More Related Videos
06:02Upper-extremity Approach for Secondary Access in Transfemoral Transcatheter Aortic Valve Implantation
Published on: August 8, 2025
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Cardiac Catheterization I: Pre-Procedure Overview