Utilizing image analysis by optical density to evaluate changes in hematoxylin and eosin staining quality after
Hannah M Benton1, Mark Butters1, Miles Brous1
1Premier Laboratory, LLC, Longmont, CO, USA.
Journal of Histotechnology
|June 23, 2025
Summary
Overusing hematoxylin and eosin (H&E) reagents decreases eosin stain intensity, but staining quality remains acceptable for pathology evaluation. This impacts digital pathology assessments.
Area of Science:
- Histology
- Pathology
- Digital Pathology
Background:
- Hematoxylin and eosin (H&E) staining is fundamental in histology.
- Digital pathology and optical density (OD) analysis offer new methods for assessing stain quality.
- Understanding factors affecting H&E staining is crucial for consistent, high-quality results.
Purpose of the Study:
- To evaluate the impact of H&E reagent overuse on staining quality.
- To assess staining precision using optical density (OD) measures.
- To analyze H&E staining characteristics across various tissue types and reagent vendors.
Main Methods:
- Combined data from two studies involving 12 tissue types (11 human, 1 porcine).
- Assessed H&E staining quality of nuclear and cytoplasmic components via OD measurements.
- Included visual inspection by a histologist and evaluation against CAP/NSH criteria.
Main Results:
- Reagent overuse led to decreased eosin stain intensity, confirmed by OD and visual inspection.
- The degree of intensity decrease varied by tissue type and reagent vendor.
- Despite overuse, H&E staining from all vendors and tissues met acceptable criteria for microscopic evaluation.
Conclusions:
- H&E reagent overuse negatively affects eosin stain intensity.
- Staining quality, while impacted, generally remains sufficient for diagnostic purposes.
- OD measures provide a quantitative method for assessing H&E staining precision in digital pathology.
Keywords:
Hematoxylin and eosinimage analysisoptical densityquality controlreagent overusereproducibilitystaining precision