In situ high-resolution cryo-EM reconstructions from CEMOVIS
Johannes Elferich1, Marek Kaminek2, Lingli Kong3
1Howard Hughes Medical Institute, University of Massachusetts Chan Medical School, Worcester, MA, USA.
Cryo-electron microscopy of vitreous sections (CEMOVIS) offers a high-resolution imaging alternative to focused-ion-beam (FIB) milling. This method provides significantly more imageable area with largely intact molecular structures, comparable to FIB-milled samples.
Area of Science:
- Structural Biology
- Cell Biology
- Microscopy Techniques
Background:
- High-resolution imaging of cells and tissues using cryo-electron microscopy (cryo-EM) requires electron-transparent samples (≤500 nm thick).
- Focused-ion-beam (FIB) milling is the standard for preparing cryo-EM samples (lamellae), but it involves material loss, limited imageable area, and ion beam damage.
- Cryo-electron microscopy of vitreous sections (CEMOVIS) presents a potential alternative to FIB milling for sample preparation.
Purpose of the Study:
- To evaluate CEMOVIS as an alternative to FIB milling for cryo-EM sample preparation.
- To assess the quality and imageable area provided by CEMOVIS.
- To demonstrate the feasibility of achieving high-resolution structural information from CEMOVIS samples.
Main Methods:
- Sample preparation using cryo-electron microscopy of vitreous sections (CEMOVIS).
- Analysis of vitreous sections, acknowledging potential damage like compression, shearing, and cracks.
- Application of two-dimensional template matching for structural reconstruction.
Main Results:
- CEMOVIS enables sectioning of samples in series, yielding significantly more imageable area compared to FIB-milled lamellae.
- Despite sectioning-induced damage, molecular structures within vitreous sections were largely intact in many regions.
- Reconstruction of the 60S ribosomal subunit from Saccharomyces cerevisiae cells at near-atomic resolution was achieved using CEMOVIS data.
Conclusions:
- CEMOVIS is a viable alternative to FIB milling for cryo-EM sample preparation, offering distinct advantages in terms of imageable area.
- The technique preserves molecular integrity sufficiently for high-resolution structural studies.
- CEMOVIS expands the possibilities for cryo-EM analysis of cellular structures, particularly for large-scale imaging studies.