In Vitro Evaluation of the Effect of Stenting on Hematological, Hemorheological and Hemodynamic Parameters, in Various Stent Configurations and Flow Conditions
Related Experiment Video
Updated: Jul 22, 2026

Experimental Investigation of Secondary Flow Structures Downstream of a Model Type IV Stent Failure in a 180° Curved Artery Test Section
Published on: July 19, 2016
D Kokkinidou1, K Kapnisis1, M Chrysostomou1
1Department of Mechanical Engineering and Materials Science and Engineering, Cyprus University of Technology, Limassol, Cyprus.
Insights
Stent implantation during percutaneous coronary intervention affects blood flow. This study found changes in red blood cell aggregation and pressure drop, particularly with specific stent configurations and higher flow rates.
Area of Science:
- Biomedical Engineering
- Cardiovascular Science
- Hemodynamics
Background:
- Percutaneous coronary intervention (PCI) is a common procedure to restore blood flow in diseased arteries.
- The impact of stent implantation on blood physiology and flow dynamics remains incompletely understood.
- Investigating stent-induced alterations in hematological, hemorheological, and hemodynamic parameters is crucial for optimizing PCI outcomes.
Purpose of the Study:
- To evaluate the effects of stent implantation on hematological, hemorheological, and hemodynamic parameters.
- To assess how different stent configurations and flow conditions influence blood flow characteristics.
- To understand the physiological consequences of stented coronary artery morphologies.
Main Methods:
- Simulated blood flow through single and overlapping stent models in straight and curved tubes.
- Utilized two flow regimes (physiological and intense) to mimic diverse conditions.
- Analyzed blood samples for hematology, red blood cell (RBC) deformability and aggregation, and viscosity; measured pressure drop across stents.
Main Results:
- Minor alterations in RBC count, hematocrit, and mean corpuscular volume observed due to prolonged stress.
- Significant increase in RBC aggregation noted, especially in the female cohort.
- Stent configuration influenced pressure drop, and blood viscosity was affected at higher flow rates.
Conclusions:
- Stent presence distinctly alters specific hemorheological and hemodynamic parameters.
- The observed changes are dependent on the experimental setup and stent configuration.
- Findings highlight the need for further research into stent design and its physiological impact.
Purpose:
Percutaneous coronary intervention is used extensively for the restoration of blood flow in diseased arteries. The influence of stent implantation on the physiology and flow of blood is an important and still not fully understood issue. The current work evaluated possible stent-induced changes in hematological, hemorheological and hemodynamic parameters.
Methods:
Experiments were performed for blood flow in single and overlapping stent configurations, in both straight and curved tube geometries, in order to reproduce various stented coronary artery morphologies. Two different flow regimes were utilized to reflect a range of physiological and more intense flow conditions. Blood samples were obtained from a healthy human population and commercially available stents were inserted in clear perfluoroalkoxy alkane tubing, connected to a syringe/syringe-pump/pressure-sensor setup. Hematological measurements, red blood cell (RBC) deformability and aggregation, and whole blood viscosity tests were performed using standard techniques. The pressure drop across the stented area was measured via an in-line pressure sensing setup.
Results:
In terms of hematology, RBC count, hematocrit, and mean corpuscular volume show a slight influence from the longer exposure to elevated stresses. Regarding hemorheology, the most profound effect was observed on RBC aggregation, with an increasing trend primarily in the female population of the study. Further, differences were found in the hemodynamics of the flow, as the pressure drop was altered according to the stent configuration. The viscosity of the blood samples is also found affected in the higher flow rate cases.
Conclusions:
The presence of the stent was found to have a distinct effect on specific hemorheological and hemodynamic parameters according to the setup and stent configuration.
More Related Videos
09:41Hemocompatibility Testing of Blood-Contacting Implants in a Flow Loop Model Mimicking Human Blood Flow
Published on: March 5, 2020
08:44An In Vitro Hemodynamic Loop Model to Investigate the Hemocytocompatibility and Host Cell Activation of Vascular Medical Devices
Published on: August 21, 2020