Related Experiment Video
Updated: Sep 18, 2025

An Inertial Measurement Unit Based Method to Estimate Hip and Knee Joint Kinematics in Team Sport Athletes on the Field
Published on: May 26, 2020
Component orientation measurements in total hip arthroplasty using an inertial measurement unit-based smart trial
Hao Tang1, Yixin Zhou2, Baojun Mai3
1Department of Orthopaedic Surgery, Beijing Jishuitan Hospital, Capital Medical University, National Orthopeadic Center of China, Beijing, 100035, China.
Background:
Intraoperative measurement of component orientation represents a basis for optimizing outcomes after total hip arthroplasty (THA). Although the use of computer navigation systems in THA has improved the accuracy of component positioning, they have not gained widespread popularity due to their complexity, time demands, and time-consuming protocols.
Methods:
We developed an Inertial Measurement Unit-based Hip Smart Trial system (IMUHST) to assist with intra-operative monitoring of hip posture. An in vitro validation experiment was conducted using a sawbones with a three-dimensional (3D) measurement model as the reference standard.
Results:
The absolute mean error, Bland-Altman analysis, and Intra-class Correlation Coefficient revealed that the accuracy and precision of this system meet the threshold for clinical application.
Conclusions:
In conclusion, this in vitro validation demonstrates that the IMUHST system provides accurate component orientation measurements while eliminating the cost and complexity of optical navigation, offering a practical solution for widespread adoption. Video Abstract.