Performance Evaluation of a Custom-Designed Contrast Media Injector in a 5-Tesla MRI Environment
Yuannan Hu1, Wenbo Sun1, Zhusha Wang1
1Department of Radiology, Zhongnan Hospital of Wuhan University, Wuchang, Wuhan 430071, China.
Abstract:
The compatibility and safety of contrast media injectors (CMIs) at ultra-high magnetic field strengths remains a critical challenge. This study aimed to investigate a custom-designed CMI powered by a ceramic motor in a newly developed 5T MRI environment, comparing it with a commercial CMI commonly used in a clinic. Three key performance aspects of the CMI were assessed in the 5T environment: translational attraction force, injection flow rates, and total injected volume. Potential imaging artifacts were checked. The custom-designed CMI demonstrated robust performance in the 5T environment, maintaining injection accuracy across all test locations and ensuring translational attraction forces remained within safe thresholds, even in the most challenging positions. Importantly, the custom-designed CMI exhibited no significant radiofrequency (RF) interference, and no imaging artifacts were observed across routine clinical sequences. In contrast, the commercial 3T CMI showed RF interference in several sensitive tests, such as the gradient echo (GRE) sequence with a 0° flip angle and frequency-based detection methods, underscoring the need for field-specific CMI designs tailored to ultra-high field environments. Further tests were performed in monkey livers and a human brain in vivo. The custom-designed CMI proved to be safe, accurate, and fully compatible with the 5T environment.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging