A feasibility study on non-invasive and non-contact jugular venous pulse measurement using 60 GHz FMCW radar
Shatabdi Das1, Hadi Afsharan2, Girish Dwivedi2,3
1School of Biomedical Engineering, The University of Sydney, Sydney, New South Wales, Australia.
Abstract:
The jugular venous pulse (JVP) reflects right atrial pressure and provides diagnostic insight into cardiovascular and pulmonary health. However, reliable assessment remains difficult due to neck adiposity, anatomical variability and suboptimal positioning. Although central venous catheterization is the gold standard, its invasive nature restricts routine or long-term use. This study introduces a non-invasive method for JVP estimation using a 60 GHz frequency-modulated continuous wave (FMCW) radar. The system captures venous pulsations at the skin surface and applies eigenbeamforming to enhance signal-to-noise ratio and pulse clarity. Radar parameters were optimized for signal fidelity and validated through morphological comparison with simultaneously recorded photoplethysmography (PPG) signals. Additionally, we compared radar-derived JVP signals with previously recorded catheterization data from a patient with early-stage heart failure to assess clinical relevance. Signal localization was successfully achieved within a direction-of-arrival (DoA) range of to , demonstrating the radar's precision. While the selected parameters consistently yielded good performance in our set-up, individual anatomical differences may require subject-specific calibration. These findings support the potential of 60 GHz FMCW radar for contactless JVP monitoring, with promising implications for early detection and remote management of heart failure.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Special considerations while measuring pulse
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: