[Quarterly and Experience-based Trends of Ferromagnetic Object Incidents in MRI Technologists: A Retrospective Study
Miho Uemura1, Yoshihiro Akatsuka1, Mitsuhiro Nakanishi1
1Division of Radiology and Nuclear Medicine, Sapporo Medical University Hospital.
Purpose:
This study was to investigate the cases of ferromagnetic objects brought into the MRI room by using incident reports.
Methods:
We included incident reports on ferromagnetic objects brought into the MRI room over the past 10 years from January 2012 to December 2021. We investigated the incidence rate of the ferromagnetic objects into the MRI room, the cause for bringing ferromagnetic objects, the number of years of MRI experience, the time of occurrence, and the names of ferromagnetic objects.
Results:
There were 248 incident reports, including 26 cases related to the ferromagnetic objects. The frequency of occurrence shows that the highest number of cases occurred within the first year of experience, accounting for approximately half of the case related to MRI staff. For newcomers, there are more incidents in the second quarter than in other quarters.
Conclusion:
The number of cases of bringing in ferromagnetic objects was higher in the second quarter as there was less experience with MRI.
Insights
New MRI staff are more likely to bring ferromagnetic objects into the MRI room, especially in their first year and during the second quarter. This highlights a need for enhanced safety training for personnel with less magnetic resonance imaging (MRI) experience.
Area of Science:
- Medical Imaging Physics
- Patient Safety
- Radiological Sciences
Context:
- Magnetic resonance imaging (MRI) rooms pose significant safety risks due to powerful magnetic fields.
- Ferromagnetic objects can cause serious accidents and injuries within the MRI environment.
- Incident reports provide valuable data for analyzing safety breaches.
Purpose:
- To investigate the incidence and causes of ferromagnetic objects being brought into MRI rooms.
- To identify trends and risk factors associated with such incidents.
- To utilize incident report data for improving MRI safety protocols.
Summary:
- Analysis of 10 years of incident reports (2012-2021) revealed 26 cases of ferromagnetic objects in MRI rooms.
- The majority of incidents involved MRI staff within their first year of experience.
- A higher frequency of incidents occurred during the second quarter, particularly among less experienced personnel.
Impact:
- Findings underscore the critical need for targeted safety education for new MRI staff.
- Recommendations for enhanced training can mitigate risks associated with ferromagnetic objects in MRI suites.
- Improved safety awareness can prevent accidents and ensure a secure MRI environment.
More Related Videos
Related Concept Videos
Magnetic Resonance Imaging
Types of Reports II: Incident or Occurrence Report
Purposes:
In the healthcare industry, reports play a crucial role in documenting incidents within an agency. The primary objective of these reports is to ensure patient safety, uphold the...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging