A Line-Source Approach for Simulating MammoWave Microwave Imaging Apparatus for Breast Lesion Detection
Navid Ghavami1, Sandra Dudley2, Mohammad Ghavami2
1UBT-Umbria Bioengineering Technologies, 06081 Perugia, Italy.
None:
Here, we propose an analytical approach to simulating MammoWave, a novel apparatus for breast cancer detection using microwave imaging. The approach is built upon the theory of cylindrical waves emitted by line sources. The sample is modelled as a cylinder with an inclusion. Our results indicate that when compared with phantom measurements, our approach gives an average relative error (between the image generated through measurement with phantoms and the image generated through the analytical simulation approach) of less than 6% when considering the full frequency band of 1-9 GHz. The procedure permits the simulation of the MammoWave imaging system loaded with multilayered eccentric cylinders; thus, it can be used to obtain an insight into MammoWave's detection capability, without having to perform either time-consuming full-wave simulations or phantom measurements.