[Accuracy Assessment of Cone-Beam CT Images for Pelvic Tumor Dose Calculation]
Bao Li1, Yongzhong Chen2, Jun Jin2
1Department of Radiotherapy, Ezhou Central Hospital, Ezhou, 436000.
Objective:
To evaluate the feasibility and accuracy of cone-beam CT (CBCT) images for radiotherapy dose calculation in pelvic tumors.
Methods:
An improved volumetric density coverage method was used to establish CT value-relative electron density (RED) curves for CBCT images. The planning CT plans were transferred to the CBCT images, and the constructed density curves were applied to calculate doses for CBCT plans while maintaining the optimization parameters unchanged. Dose calculation deviations between the two plans were analyzed.
Results:
The mean differences in dosimetric parameters for the target volume and organs at risk (OAR) between the two plans were less than 1% and 1.5%, respectively. The target conformity index (CI), homogeneity index (HI), and gamma passing rates were highly consistent, with no statistically significant differences.
Conclusion:
CBCT images corrected by this method can be used for dose calculation in pelvic tumor radiotherapy planning.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography