Flexible and navigable suction access sheaths: what size stone particles can be cleared?
Richard Menzies-Wilson1, Jessica Williams2, Koushikk Ayyappan1
1Nuffield Department of Surgery, University of Oxford, Oxford, UK.
Objectives:
To perform benchtop experiments on flexible and navigable suction (FANS) ureteric access sheaths (UASs) to determine the clearance capabilities for various stone sizes when paired with different-sized ureteroscopes.
Methods:
Quartz stones were used as a surrogate for renal stones. Stone samples were created to simulate the range of particle sizes produced by 'dusting' a 1-cm calcium oxalate monohydrate stone. The stone mixture was introduced into an open vial at a 30° angle in aliquots every 3 min over the course of 30 min, mimicking their gradual production by lasertripsy. This benchtop model focused on the effects of geometry (between FANS access sheath and ureteroscope) and suction, excluding complex fluid flow. An 11/13-F ClearPetra® FANS UAS with an indwelling ureteroscope was free to move in the vial and targeted visible stones. The vial was intermittently filled to 40 mL with water and emptied using 200-mmHg suction. The experiment had two arms: (1) a permanently indwelling ureteroscope in the FANS UAS to mimic continuous lasering with aspiration and (2) intermittent withdrawal of the ureteroscope to mimic pausing lasering to clear larger fragments through the empty FANS sheath lumen. Three ureteroscope sizes were used: 9.5-F (Lithovue™), 7.5 F (PUSEN) and 6.3 F (Hugemed). The experiment was performed four times for each ureteroscope. [Correction added on 4 July 2025, after first online publication: In the preceding sentence, "three" has been corrected to "four" in this version.] RESULTS: With a permanently indwelling 9.5-F ureteroscope, the FANS UAS cleared 64% of the overall stone mass but became blocked after an average of 21 min. Intermittent withdrawal of the ureteroscope cleared all stones. With indwelling 7.5-F and 6.3-F ureteroscopes, the FANS UAS did not become blocked and completely cleared stones of up to 500 μm and 2 mm, respectively.
Conclusions:
Without withdrawing the ureteroscope (potentially allowing continuous lasertripsy), it is possible to continuously aspirate small fragments alongside a 7.5-F and a 6.3-F ureteroscope in an 11/13-F FANS UAS. Intermittent withdrawal of a ureteroscope from the FANS UAS allows complete stone clearance.
Related Concept Videos
Quarrying of Stone
One common method involves using a diamond belt saw to cut large blocks from the quarry face. These blocks can be about 50 feet long and 12 feet high. After the initial vertical cut, drilling is performed at the base of the...
Urinary Tract Calculi VI: Surgical Management
Maximum Size of Aggregate
Sieve Analysis and Grading Curves
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Pumped Concrete
For direct-acting pumps, the concrete enters the pump via the inlet valve under the action of gravity and suction created by the movement of the piston. This concrete is then forced into the pipeline and out through the outlet valve by the forward movement...