Evaluation of Six Commercial and Noncommercial Colistin Resistance Diagnostics

Tumisho Mmatumelo Seipei Leshaba1, Masego Mmatli1, Nontombi Marylucy Mbelle1

  • 1Department of Medical Microbiology, School of Medicine, University of Pretoria, Pretoria, South Africa.

Microbiologyopen
|June 30, 2025
PubMed

Insights

Colistin resistance is rising globally. Vitek 2 and Sensititre offer the best performance for colistin minimum inhibitory concentration (MIC) testing, while Rapid NP provides a quick screen.

Area of Science:

  • Clinical Microbiology
  • Antimicrobial Resistance
  • Diagnostic Test Evaluation

Background:

  • Colistin is a critical last-resort antibiotic for treating multidrug-resistant Gram-negative infections.
  • Rising global colistin resistance necessitates accurate and efficient diagnostic methods.
  • Several commercial and in-house diagnostic tests are available for detecting colistin resistance.

Purpose of the Study:

  • To evaluate and compare the performance of six colistin resistance diagnostics against a gold standard.
  • To determine the sensitivity, specificity, categorical agreement, and error rates of each test.
  • To guide the selection of appropriate diagnostic tools for clinical laboratories.

Main Methods:

  • Broth microdilution (BMD) according to ISO standards was used as the reference method.
  • 142 Gram-negative isolates (Enterobacterales, Acinetobacter baumannii, Pseudomonas aeruginosa) were tested.
  • Multiplex PCR screening for mcr genes (mcr-1 to mcr-5) was performed.

Main Results:

  • Vitek 2 and Sensititre demonstrated the highest categorical agreement (≥98%) and no very major errors.
  • ComASP showed excellent performance for Acinetobacter baumannii but lower agreement for Pseudomonas aeruginosa.
  • CHROMagar COL-APSE had low specificity (69%) and high major errors in Enterobacterales; Rapid NP test showed a high very major error rate (7.84%).

Conclusions:

  • Vitek 2 and Sensititre are optimal for colistin MIC testing, with ComASP as a cost-effective alternative.
  • The Rapid NP test offers rapid screening but requires confirmatory MIC testing.
  • CHROMagar COL-APSE requires careful inoculum control to minimize false resistance detection.