High-pitch photon-counting detector computed tomography angiography of the coronary arteries: Qualitative and quantitative evaluation of monoenergetic image reconstructions
Related Experiment Video
Updated: Sep 17, 2025

High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Andreas Strassl1,2, Francesco Lauriero1,3, Maria Alejandra Rueda1,4
1Division of Cardiovascular and Interventional Radiology, Department of Biomedical Imaging and Image-guided Therapy, Medical University of Vienna, Währinger Gürtel 18-20, Vienna 1090, Austria.
Background:
Dual-source photon-counting detector computed tomography (PCDCT) offers the opportunity to perform cardiac examinations within one beat and simultaneously the acquisition of spectral information. This study, evaluated subjective and objective image quality of virtual monoenergetic image (VMI) reconstructions using data from a first-generation, dual-source PCDCT scanner, operated in high-pitch scanning mode.
Methods:
We retrospectively included 30 patients who underwent a clinically indicated CTA of the coronary arteries. VMI were reconstructed at five different energy levels. Subjective image quality was assessed by three radiologists according to a four-point Likert scale for four different quality features. To evaluate objective image quality, SNR and CNR were calculated via ROIs placed in the aorta, coronary arteries, myocardium, pectoral muscle, and epicardial fat.
Results:
VMI at 40, 50, 60, and 70 keV showed equal mean scores (4/4) for subjective vascular contrast, followed by 80 keV reconstructions with a mean score of 3/4. The 40 keV reconstruction yielded the lowest range (3-4) in Likert scores and highest percentage of reader agreement (80 %). Minor differences in subjective image noise, sharpness, and plaque visualization were observed with positive trends toward higher keV levels. SNR and CNR were superior for 40 keV, with a mean of 34.8 ± 1.7HU and 45.4 ± 2.7HU, respectively. Mean applied contrast volume was 65 ml, resulting in a mean CT value of 1150HU for 40 keV VMI.
Conclusion:
First-generation PCDCT-derived VMI at 40 and 50 keV offer satisfying subjective and objective image quality, even when acquired in high-pitch scanning mode.
More Related Videos
06:57Author Spotlight: Advancing Cardiovascular Imaging - Introducing the Spatially Weighted Calcium Score for Early Disease Detection
Published on: September 22, 2023
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...