Sex Differences in BP From an Automated Oscillometric Compared With Manual Device
Tan V Bui1, James E Sharman1, Niamh Chapman1,2
1Menzies Institute for Medical Research, University of Tasmania, Hobart, Australia (T.V.B., J.E.S., N.C., M.G.S., S.L.G., D.S.P.).
Background:
Automated blood pressure (BP) devices may be less accurate in females than males, but this requires further investigation. This study aimed to determine sex differences in automated BP, measured with a single brand and model of device, compared with manual BP, with a focus on cuff sizes and associations with measures of adiposity.
Methods:
Automated (Omron HEM-907XL) and manual BP were taken sequentially in a random order among a subsample of participants attending the US National Health and Nutrition Examination Survey, 2017 to 2018. Anthropometry and dual-energy x-ray absorptiometry were used to record body size and composition. Analyses, including multivariable regression to determine sex differences in BP, by cuff size, followed complex survey statistical principles.
Results:
A total of 3735 participants (49.0% female [95% CI, 46.4-51.6], 45 years [43-46]) were included. In females, automated systolic BP (SBP) incrementally underestimated manual SBP across larger cuffs up to extra-large (-6.4 mm Hg [-8.0 to -4.9]). In males, automated SBP underestimated manual SBP only with extra-large cuffs (-2.4 mm Hg [95% CI-3.9 to -0.9]). Underestimation by automated SBP with extra-large cuffs was independently associated with all measures of body size indicative of increased adiposity in both females and males. Hypertension classification from automated and manual SBP had moderate agreement for adult/large cuffs (weighted kappa range 0.66-0.79) but weak agreement for extra-large cuffs (0.55-0.58) for females and males.
Conclusions:
The automated device used in this study underestimated manual SBP at larger cuff sizes, which was associated with indices of adiposity. Poorer accuracy of automated BP in larger cuff sizes could contribute to inequitable BP-related health care for females and males and requires further investigation.
More Related Videos
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure