Sex Differences in BP From an Automated Oscillometric Compared With Manual Device.
Tan V Bui1, James E Sharman1, Niamh Chapman1,2
1Menzies Institute for Medical Research, University of Tasmania, Hobart, Australia (T.V.B., J.E.S., N.C., M.G.S., S.L.G., D.S.P.).
Automated blood pressure devices underestimated readings more in females than males, especially with larger cuffs. This inaccuracy, linked to adiposity, may impact equitable healthcare.
Area of Science:
- Cardiovascular Health
- Medical Device Accuracy
- Public Health
Background:
- Automated blood pressure (BP) devices may exhibit differential accuracy between sexes.
- Previous research indicates potential disparities, necessitating focused investigation.
- Cuff size and body composition are critical factors influencing BP measurement accuracy.
Purpose of the Study:
- To investigate sex differences in automated BP device accuracy compared to manual BP.
- To examine the impact of varying cuff sizes on automated BP measurements.
- To assess the association between adiposity measures and automated BP accuracy.
Main Methods:
- Utilized data from the US National Health and Nutrition Examination Survey (2017-2018).
- Compared automated (Omron HEM-907XL) and manual BP measurements in a random subsample.
- Employed anthropometry and dual-energy x-ray absorptiometry for body size and composition analysis.
Main Results:
- Females showed greater underestimation of systolic BP (SBP) with larger cuffs compared to males.
- Automated SBP underestimation with extra-large cuffs correlated with increased adiposity in both sexes.
- Agreement between automated and manual BP for hypertension classification was moderate with standard cuffs but weak with extra-large cuffs.
Conclusions:
- Automated BP devices, particularly with larger cuffs, can underestimate SBP, with accuracy linked to adiposity.
- Discrepancies in automated BP accuracy may contribute to healthcare inequities for females and males.
- Further research is essential to address the implications of automated BP device accuracy on patient care.
More Related Videos
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure