Summary
Physical therapy students
Area of Science:
- Biomechanics
- Physical Therapy
- Medical Measurement
Background:
- Goniometry is a common method for measuring joint angles.
- Accuracy in goniometry is crucial for effective physical therapy assessment.
- Potential sources of error in goniometry require investigation.
Purpose of the Study:
- To assess the accuracy of goniometry in measuring elbow joint angles.
- To identify potential sources of error in goniometric measurements.
- To compare goniometric accuracy with a photographic reference standard.
Main Methods:
- Forty-six physical therapy students performed elbow goniometry using plastic or steel goniometers.
- Three measurement protocols were used: ALIGN, ASSIGN, and PROM (passive range of motion).
- Measurements were compared against a photographic reference standard for accuracy.
Main Results:
- Goniometric accuracy (s +/- 2.4-3.4 degrees) was lower than photographic accuracy (s +/- 0.7-1.1 degrees).
- Goniometer alignment, landmark identification, and manual force during PROM contributed to measurement error.
- Most goniometric measurements showed statistically significant differences from photographic measurements (p < .05).
Conclusions:
- Small systematic errors in alignment and landmark identification may explain discrepancies.
- Standardized methods enable even inexperienced raters to use goniometers accurately for elbow position measurement.