Related Experiment Video
Updated: Sep 17, 2025

Multi-Modal Home Sleep Monitoring in Older Adults
Published on: January 26, 2019
Diagnostic Accuracy of the Alice NightOne Single-Belt Monitor for Obstructive Sleep Apnea and Reliability of Wireless
MaoHuan Peng1, YuanYuan Zhang2, Rui Zhao3
1Department of Sleep Medicine, Peking University People's Hospital, Beijing, People's Republic of China.
Objective:
To validate the performance of a single-belt type 3 portable monitor Alice NightOne for diagnosis of obstructive sleep apnea and its reliability in remote data transmission.
Methods:
Our study included two parts: in-laboratory Alice NightOne (ANO) monitoring (ANOlab) and home sleep apnea testing (HSAT) using ANO (ANOhome). For ANOlab, the participants underwent polysomnography (PSG) with ANOlab simultaneously. For ANOhome, the participants completed unattended overnight ANOhome out of sleep center and PSG was performed in another night. The ANO recordings were transmitted to cloud database wirelessly in addition to traditional wire transmission. Message digest-5 (MD5) algorithm was utilized to verify the integrity of the cloud data.
Results:
Ninety-one ANOlab and 170 ANOhome recordings were analyzed. Both the respiratory event index (REI) on ANOlab and that on ANOhome were lower than the corresponding apnea-hypopnea index (AHI) on PSG (24.9 ± 20.5 events/h vs 31.6 ± 25.0 events/h, and 26.7 ± 17.0 events/h vs 35.3 ± 21.2 events/h respectively, P < 0.001). Bland-Altman analysis of REI on ANOlab versus AHI on PSG showed a mean difference (95% confidence interval) of -6.7 (-8.4, -4.9) events/h. For REI on ANOhome versus AHI on PSG, the difference is -8.0 (-9.9, -6.0) events/h. With threshold of REI ≥ 5 events/h for OSA diagnosis, ANOhome had 98.8% sensitivity, 90.0% specificity, 99.4% positive predictive value. The MD5 algorithm verified the identity between uploaded cloud data and original data.
Conclusion:
With single thoracoabdominal belt, Alice NightOne can help diagnosis of obstructive sleep apnea with good sensitivity and specificity, though it may underestimate AHI. Furthermore, it provides reliable support based on solid data teletransmission and scoring synchronization, which may increase the ability of diagnosis and management of OSA through telemedicine.