A feasibility study of a computational modeling system for performance evaluation and development of ultrasound
Marjan Bakhtiari-Nejad1, Nima Akhlaghi2, Reza Zahiri Azar3
1Center for Devices and Radiological Health, U.S. Food and Drug Administration, Silver Spring, MD, United States of America.
Abstract:
Ultrasound strain elastography (USE) is an imaging technology that enables us to detect changes in tissue stiffness resulting from cancer and other diseases. The objective of this study is to computationally model the application of USE for breast lesion characterization. We develop a well-defined simulation pipeline using open-source software to createin silicoUSE phantoms with one and two stiff targets. First, we use FreeCAD software for tissue 3D modeling and Gmsh software for finite element (FE) meshes. Second, we place randomly positioned point scatterers within the meshed models to form pre-deformation virtual ultrasound phantoms. Then, a simulated ultrasound transducer is used to compress and deform tissue in FE simulations using FEBio software to create a post-deformation virtual ultrasound phantom. Third, we use the k-Wave acoustics toolbox to generate pre- and post-deformation ultrasound echo signals and B-mode images. Finally, we estimate axial and lateral displacements using a speckle tracking method, and strain elastograms, using a least-squares method. Displacements from the USE simulation pipeline and phantom experiments were compared against true FEBio-simulated displacements for accuracy. We have also quantitatively compared the resultant strain elastograms obtained from FEBio simulations, USE simulation pipeline, and phantom experiments. Finally, model validation is performed by comparing the performance of the USE software platform and physical phantom experiments for a range of compression values (0.5%-5% axial strain). The results confirm the use of the well-validated USE simulation pipeline as a robust non-clinical assessment tool for USE system development.
More Related Videos
14:14Quantification of Strain in a Porcine Model of Skin Expansion Using Multi-View Stereo and Isogeometric Kinematics
Published on: April 16, 2017
06:18Intravascular Ultrasound Image-Based Finite Element Modeling Approach for Quantifying In Vivo Mechanical Properties of Human Coronary Artery
Published on: December 6, 2024